Bispectral index, spectral edge frequency 95%, and median frequency recorded for various concentrations of isoflurane and sevoflurane in pigs.

作者信息

Martín-Cancho María F, Lima Juan R, Luis Laura, Crisóstomo Verónica, Ezquerra Luis J, Carrasco María S, Usón-Gargallo Jesús

机构信息

Minimally Invasive Surgery Centre, Avda/Universidad s/n 10071 Cáceres, Spain.

出版信息

Am J Vet Res. 2003 Jul;64(7):866-73. doi: 10.2460/ajvr.2003.64.866.

Abstract

OBJECTIVE

To evaluate bispectral index (BIS), spectral edge frequency 95% (SEF), and median frequency (MED) in relation to a visual analogue scale (VAS) as indicators of anesthetic depth for various concentrations of sevoflurane and isoflurane in pigs.

ANIMALS

32 pigs.

PROCEDURE

Pigs were randomly allocated to 8 groups (4 pigs/group). An electroencephalogram (EEG) was recorded in each conscious pig. Pigs were then anesthetized by use of sevoflurane (n = 16) or isoflurane (16). Agents were administered in oxygen at minimum alveolar concentrations (MACs) of 1, 1.25, 1.5, and 1.75 MAC in a randomized order. End-tidal sevoflurane and isoflurane concentrations were maintained for 30 minutes, after which an EEG was recorded for 5 minutes; BIS, SEF, and MED were then calculated. Anesthetic depth was evaluated by use of the VAS. Cardiovascular and EEG responses to nociceptive stimuli were evaluated for each anesthetic agent.

RESULTS

BIS decreased significantly for the various concentrations of each anesthetic. At equivalent MACs, BIS values were significantly higher during sevoflurane-induced anesthesia than during isoflurane-induced anesthesia. Values of MED and SEF decreased significantly from basal values to 1 MAC of sevoflurane and isoflurane. For both agents, there was good correlation between VAS scores and BIS values and between VAS scores and SEF values.

CONCLUSIONS AND CLINICAL RELEVANCE

BIS was useful for predicting changes in anesthetic depth at clinical dosages of inhalant anesthetics. Values of BIS, SEF, and MED were significantly higher during anesthesia induced by administration of sevoflurane than during anesthesia induced by administration of isoflurance at equivalent MACs.

摘要

目的

评估双谱指数(BIS)、频谱边缘频率95%(SEF)和中位频率(MED)与视觉模拟评分法(VAS)的关系,以作为猪体内不同浓度七氟醚和异氟醚麻醉深度的指标。

动物

32头猪。

方法

猪被随机分为8组(每组4头)。记录每头清醒猪的脑电图(EEG)。然后使用七氟醚(n = 16)或异氟醚(16)对猪进行麻醉。药物以随机顺序在氧气中以1、1.25、1.5和1.75最低肺泡浓度(MAC)给药。维持呼气末七氟醚和异氟醚浓度30分钟,之后记录5分钟EEG;然后计算BIS、SEF和MED。使用VAS评估麻醉深度。评估每种麻醉剂对伤害性刺激的心血管和EEG反应。

结果

每种麻醉剂不同浓度下BIS均显著降低。在等效MAC时,七氟醚诱导麻醉期间的BIS值显著高于异氟醚诱导麻醉期间的BIS值。七氟醚和异氟醚从基础值到1 MAC时,MED和SEF值均显著降低。对于两种药物,VAS评分与BIS值之间以及VAS评分与SEF值之间均存在良好相关性。

结论及临床意义

BIS可用于预测吸入麻醉药临床剂量下麻醉深度的变化。在等效MAC时,七氟醚诱导麻醉期间的BIS、SEF和MED值显著高于异氟醚诱导麻醉期间的这些值。