Kagemann Larry, Harris Alon, Jonescu-Cuypers Christian, Kumar Rishi, Sheets Clinton, Rechtman Ehud, Migliardi Renata, Garzozi Hanna J
Department of Ophthalmology, Indiana University School of Medicine, Indianapolis, Indiana 46202, USA.
Ophthalmic Surg Lasers Imaging. 2003 Jul-Aug;34(4):342-7.
To compare measurements of a new retinal blood flow device with central retinal artery blood velocity.
One randomly selected eye from each of 13 subjects was examined. Blood flow was measured by laser blood flowmeter and velocity by color Doppler imaging. The correlation between measurements was analyzed using regression analysis where a P value of less than .05 and coefficient of regression values of greater than .5 were considered significant.
The laser blood flowmeter produced vessel diameter measurements of 90.1 +/- 18.7 microm (mean +/- standard deviation), velocity of 19.7 +/- 8.06 mm/sec, and flow measurements of 4.24 +/- 2.41 microL/min. Central retinal artery peak systolic velocity, end diastolic velocity, and mean velocity correlated significantly with laser blood flowmeter velocity (P = .01, r = .66; P < .01, r = .77; and P = .003, r = .76, respectively) and flow (P = .01, r = .71; P = .03, r = .6; and P = .01, r = .69, respectively).
The laser blood flowmeter produces retinal artery flow and velocity measurements that correlate with central retinal artery peak systolic velocity end diastolic velocity, and mean velocity measurements. Further validation of the laser blood flowmeter's accuracy in measuring real flow warranted and likely requires more invasive in vivo studies (in animal models). However, this study supports the ability of the laser blood flowmeter to measure retinal blood flow.
比较一种新型视网膜血流检测设备的测量结果与视网膜中央动脉血流速度。
对13名受试者每人随机选取一只眼睛进行检查。使用激光血流仪测量血流,用彩色多普勒成像测量血流速度。采用回归分析来分析测量结果之间的相关性,P值小于0.05且回归系数值大于0.5被视为具有显著性。
激光血流仪测得血管直径为90.1±18.7微米(平均值±标准差),血流速度为19.7±8.06毫米/秒,血流量为4.24±2.41微升/分钟。视网膜中央动脉的收缩期峰值速度、舒张末期速度和平均速度与激光血流仪的血流速度(分别为P = 0.01,r = 0.66;P < 0.01,r = 0.77;P = 0.003,r = 0.76)和血流量(分别为P = 0.01,r = 0.71;P = 0.03,r = 0.6;P = 0.01,r = 0.69)显著相关。
激光血流仪所测得的视网膜动脉血流量和血流速度与视网膜中央动脉的收缩期峰值速度、舒张末期速度及平均速度测量值相关。激光血流仪在测量实际血流量方面的准确性有待进一步验证,可能需要在动物模型中进行更具侵入性的体内研究。然而,本研究支持激光血流仪测量视网膜血流量的能力。