Załuska Wojciech T, Małecka Teresa, Swatowski Andrzej, Ksiazek Andrzej
Department of Nephrology, Medical University of Lublin.
Ann Univ Mariae Curie Sklodowska Med. 2002;57(2):337-41.
The bioimpedance technique is postulated as not expensive and non-invasive method for measurement of body water compartments in humans. However, the accuracy of the whole body bioimpedance technique has been questioned by several authors because of nonhomogenic nature of the human body. In the present study we used the whole body bioimpedance technique in comparison to segmental measurement from the arm and leg for assessment of extracellular water compartment. The extracellular water compartment value as measured by the whole body technique significantly correlated with extracellular water compartment value measured from segmental measurement in the leg. We postulate practical use of segmental measurement from the leg as an indicator of body hydration.
生物电阻抗技术被认为是一种用于测量人体水代谢区间的低成本、非侵入性方法。然而,由于人体的非均匀性,全身生物电阻抗技术的准确性受到了几位作者的质疑。在本研究中,我们使用全身生物电阻抗技术,并与手臂和腿部的分段测量进行比较,以评估细胞外水代谢区间。通过全身技术测量的细胞外水代谢区间值与通过腿部分段测量得到的细胞外水代谢区间值显著相关。我们推测,将腿部的分段测量实际应用作为身体水合作用的指标。