Nykänen Kirsi, Siltanen Samuli
Instrumentarium Corp., Imaging Div., PO Box 20, Tuusula, 04301, Finland.
Med Phys. 2003 Jul;30(7):1864-73. doi: 10.1118/1.1584160.
Effects of x-ray scattering on full-field digital mammography are analyzed with the scattering model of Seibert and Boone [Med. Phys. 15, 567-575 (1988)]. A new method is introduced for the estimation of model parameters from measurements. It is shown that with breasts thinner than a certain threshold, removing the anti-scatter grid leads to an improved contrast-to-noise ratio with a smaller patient dose. A fast approximate algorithm is presented for determining the scattered field in a gridless digital mammogram.