Toroi P, Nieminen M T, Tenkanen-Rautakoski P, Varjonen M
STUK-Radiation and Nuclear Safety Authority, 00881 Helsinki, Finland.
Phys Med Biol. 2009 Jun 21;54(12):3865-79. doi: 10.1088/0031-9155/54/12/017. Epub 2009 Jun 2.
The pixels in digital detectors used in x-ray imaging can be considered as small dose meters. In mammography, the image detector area is commonly only partly covered by tissue and the incident air kerma could be monitored by using pixel values from the detector elements in the unattenuated primary beam area. The measured air kerma could be used for patient dose estimation. The air kerma and radiation quality dependence of the response of pixel values within a large exposure range for two direct digital detector types was experimentally investigated for this purpose. Using these calibration results, air kerma was measured from clinical images and compared with tube output calculation. The differences between the two methods were typically below 2% and measurements based on pixel values can therefore be recommended.
X射线成像中使用的数字探测器中的像素可被视为小型剂量计。在乳腺摄影中,图像探测器区域通常仅部分被组织覆盖,并且可以通过使用未衰减的原射线束区域中探测器元件的像素值来监测入射空气比释动能。所测量的空气比释动能可用于患者剂量估计。为此,通过实验研究了两种直接数字探测器类型在大曝光范围内像素值响应的空气比释动能和辐射质量依赖性。利用这些校准结果,从临床图像中测量空气比释动能,并与管输出计算结果进行比较。两种方法之间的差异通常低于2%,因此可以推荐基于像素值的测量方法。