Instrument-independent flux units for laser Doppler perfusion monitoring assessed in a multi-device study on the renal cortex.

作者信息

Petoukhova A L, Steenbergen W, Morales F, Graaff R, de Jong E D, Elstrodt J M, de Mul F F M, Rakhorst G

机构信息

Faculty of Applied Physics, Biophysical Techniques Group, University of Twente, P.O. Box 217, 7500 AE Enschede, The Netherlands.

出版信息

Microvasc Res. 2003 Sep;66(2):83-90. doi: 10.1016/s0026-2862(03)00036-0.

Abstract

To investigate the feasibility of instrument-independent perfusion units for laser Doppler flowmetry, a comparison was performed of two commercial fiberoptic laser Doppler perfusion monitors measuring the same flux situation for two different types of probes. In vivo measurements were performed on the cortex of pig's kidney, with an ultrasonic arterial flow meter as reference. The flow was mainly varied by internal arterial constriction using a balloon catheter. For each probe, instruments are compared in terms of the ratio of laser Doppler flux and arterial flow. For a given probe, the flux-to-flow ratios of the two instruments show a linear mutual relationship for a wide variety of arterial flows and laser Doppler fluxes. In vitro measurements were performed on an aqueous suspension of polystyrene microspheres. For the probe with interfiber distance 500 microm the ratio of the in vivo fluxes appears to agree within 16% to the value found in vitro, while for the 250-microm probe a difference of 28% was found. For a wide range of fluxes, the in vivo flux values of one instrument can be translated into flux values for the other instrument, in spite of the instrumental differences. This enables the user to render experimental results independent of the specific instrument, thus facilitating multi-center studies.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验