Experiments on extracted teeth into the validity of using laser Doppler techniques for recording pulpal blood flow.

作者信息

Vongsavan N, Matthews B

机构信息

Department of Physiology, University of Bristol, University Walk, U.K.

出版信息

Arch Oral Biol. 1993 May;38(5):431-9. doi: 10.1016/0003-9969(93)90215-8.

Abstract

The performances of two laser Doppler flow meters (Periflux model PF3 and Moor Blood Flow Monitor model MBF3D) were investigated. Recordings were made while diluted blood was pumped at different rates through the pulp cavities of extracted human or pig teeth. The probe of each instrument was fixed to the enamel surface 2 mm from the original position of the gingival margin. Both instruments performed similarly, although the Moor gave the better signal/noise ratios. Both were capable of detecting flow in the core of the pulp as well as the superficial part closest to the probe tip. With either instrument, and with any one dilution of blood (range 0.5-45% v/v red cells), there was a near-linear relationship between the blood-flow signal and the flow rate of blood through the tooth. However, when data obtained with different red-cell concentrations were compared, a good correlation between the blood-flow signal and red-cell flux (product of concentration and mean velocity) was obtained only with red-cell concentrations up to 1% v/v. Therefore these instruments would give an accurate indication of blood flow changes only under conditions in which either this value was not exceeded or if the red-cell volume fraction remained constant; neither of which can be assumed to apply when recording from teeth in situ. The signals representing the concentration of moving blood cells were unreliable in both instruments. It is concluded that the information provided by these laser Doppler flow meters can be ambiguous and must be interpreted with care.

摘要

对两种激光多普勒血流仪(Periflux型号PF3和Moor血流监测仪型号MBF3D)的性能进行了研究。在将稀释血液以不同速率泵入拔除的人牙或猪牙的髓腔时进行记录。每种仪器的探头固定在距牙龈边缘原始位置2毫米的牙釉质表面。两种仪器的表现相似,尽管Moor仪器的信噪比更好。两者都能够检测到牙髓核心以及最靠近探头尖端的表层部分的血流。使用任何一种仪器,以及任何一种血液稀释度(红细胞体积分数范围为0.5 - 45% v/v),血流信号与通过牙齿的血流速率之间都存在近似线性关系。然而,当比较不同红细胞浓度下获得的数据时,仅在红细胞浓度高达1% v/v时,血流信号与红细胞通量(浓度与平均速度的乘积)之间才具有良好的相关性。因此,这些仪器仅在不超过该值或红细胞体积分数保持恒定的条件下才能准确指示血流变化;而在对原位牙齿进行记录时,这两种情况都不能假定成立。两种仪器中代表移动血细胞浓度的信号都不可靠。结论是,这些激光多普勒血流仪提供的信息可能含糊不清,必须谨慎解读。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验