[X线放大摄影技术中辐射散射的减少(作者译)]

[Reduction in radiation scatter during radiographic magnification techniques (author's transl)].

作者信息

Stargardt A, Angerstein W

出版信息

Rofo. 1975 Oct;123(4):364-9. doi: 10.1055/s-0029-1230216.

Abstract

The reduction in scatter during magnification techniques is compared with the effect of grids. The results of extensive measurements concerning scatter and distance in a water phantom are shown in 24 diagrams. The various factors were altered as follows: distance 0-1.5 m., phantom thickness 0-27.5 cm., area 5 cm. x 5 cm. - 30 cm. x 30 cm. and kilovoltage 50-150 kv. The amount of scatter immediately behind the phantom is compared with that through a grid by observing image quality for given amounts of scatter. This "grid-equivalent" image quality can be used to determine the circumstances under which a magnification technique can be employed without using a grid.

摘要

将放大技术中散射的减少与滤线栅的效果进行了比较。在水模体中关于散射和距离的大量测量结果显示在24张图表中。各种因素的改变如下:距离0 - 1.5米,模体厚度0 - 27.5厘米,面积5厘米×5厘米 - 30厘米×30厘米,以及千伏值50 - 150千伏。通过观察给定散射量下的图像质量,将模体后方紧邻处的散射量与通过滤线栅的散射量进行比较。这种“滤线栅等效”的图像质量可用于确定在不使用滤线栅的情况下采用放大技术的情形。

相似文献

5
[Use of multi-grid screens in intensive care units].
Aktuelle Radiol. 1995 Sep;5(5):293-6.
6
Image quality in mammography with special reference to anti-scatter grids and the magnification technique.
Acta Radiol Diagn (Stockh). 1986 Jul-Aug;27(4):467-79. doi: 10.1177/028418518602700420.
8
Grid versus air gap. A comparison of cephalometric techniques.格栅与气隙。头影测量技术的比较。
Oral Surg Oral Med Oral Pathol. 1994 Jan;77(1):86-9. doi: 10.1016/s0030-4220(06)80113-1.
10
Space - the unknown accessory.太空——未知的附属物。
Radiol Technol. 1981 May-Jun;52(6):597-604.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验