Ding J, Liao Z, Wang L, Yang X
Foshan College of Science and Technology, 528000 Foshan.
Guang Pu Xue Yu Guang Pu Fen Xi. 2001 Aug;21(4):552-4.
A method for the determination of trace cadmium in chinese medicine Loulu by graphite furnace atomic absorption spectrometry (GFAAS) has been described. The matrix modifier and its proper concentration, the influence factors including ashing and atomization temperature, atomization time and coexisting ions were investigated. Under the optimum experimental conditions, the detection limit and the relative standard deviation of the method were 0.06 ng.mL-1 and 2.43% (n = 11), respectively. The recovery was between 97% and 105%. The method is simple, rapid and accurate, and can eliminate the matrix effect. It has been applied to the direct determination of trace cadmium in chinese medicine Loulu with satisfactory results.
本文描述了一种用石墨炉原子吸收光谱法(GFAAS)测定中药漏芦中痕量镉的方法。研究了基体改进剂及其合适浓度,考察了包括灰化温度、原子化温度、原子化时间和共存离子等影响因素。在最佳实验条件下,该方法的检出限和相对标准偏差分别为0.06 ng·mL-1和2.43%(n = 11)。回收率在97%至105%之间。该方法简便、快速、准确,可消除基体效应。已应用于直接测定中药漏芦中的痕量镉,结果令人满意。