Analysis of artemisinin by a packed-column supercritical fluid chromatography-atmospheric pressure chemical ionisation mass spectrometry technique.

作者信息

Dost Kenan, Davidson George

机构信息

Department of Chemistry, University of Celal Bayar, Manisa, Turkey.

出版信息

Analyst. 2003 Aug;128(8):1037-42. doi: 10.1039/b301342a.

Abstract

A packed-column supercritical fluid chromatography-atmospheric pressure chemical ionisation mass spectrometry method was studied for the determination of artemisinin from Artemisia annua L. extracts. The technique does not require any kind of derivatisation prior to the analysis. All samples were simply dissolved in methanol and injected into the mobile phase. Detection was achieved by using mass spectrometry with atmospheric pressure chemical ionisation. The ionisation technique is relatively soft and provides protonated molecular ion and informative structural fragmentation for the compound. Benzophenone was used as a chromatographic standard for the determination of the analytical reproducibility. The supercritical carbon dioxide mobile phase used in the system was modified by 10% methanol. The average absolute retention time was 3.54 min with a standard deviation of 0.017 min and a relative standard deviation of 0.4% with respect to benzophenone for the procedure. The correlation coefficient was 0.998 and detection limit 370 pg on column.