Scheurlen M, Waller C, Hartmann F
Department of Gastroenterology, Medizinische Universitätsklinik, Tübingen, Germany.
J Clin Ultrasound. 1992 Jan;20(1):19-26. doi: 10.1002/jcu.1870200105.
Image-directed Doppler measurement of superior mesenteric artery blood flow in volunteers was validated in two artificial models with either a venous or an arterial flow profile. In the "venous" model, the Doppler device overestimated the real flow velocity by 86 +/- 7%, and in the "arterial" model by 24 +/- 9%. The areas under the time-frequency curves (AUCs) from the arterial model were measured by planimetry. A correction factor of 1.47 between flow velocities calculated after planimetry and real flow was established. Correction for this factor resulted in a flow velocity of 19.5 +/- 4.7 cm/s and a blood flow rate of 377 +/- 166 ml/min in the volunteers.
在具有静脉或动脉血流剖面的两个人造模型中,对志愿者进行了图像引导的肠系膜上动脉血流多普勒测量验证。在“静脉”模型中,多普勒设备将实际流速高估了86±7%,在“动脉”模型中高估了24±9%。通过平面测量法测量动脉模型的时频曲线下面积(AUCs)。在平面测量法计算的流速与实际流速之间建立了1.47的校正因子。对该因子进行校正后,志愿者的流速为19.5±4.7厘米/秒,血流量为377±166毫升/分钟。