肠系膜血流速度随声束入射角度的变化。

Mesenteric flow velocity variations as a function of angle of insonation.

作者信息

Rizzo R J, Sandager G, Astleford P, Payne K, Peterson-Kennedy L, Flinn W R, Yao J S

机构信息

Department of Surgery, Northwestern University Medical School, Chicago, IL.

出版信息

J Vasc Surg. 1990 May;11(5):688-94. doi: 10.1067/mva.1990.19707.

Abstract

This study was designed to quantitate variations in duplex ultrasound arterial flow velocities (cm/sec) in the common carotid artery and the superior mesenteric artery that were produced by changes in the angle of pulsed Doppler insonation. Duplex scanning was used to measure peak systolic flow velocity and mean velocity at angles from 30 degrees to 80 degrees; individual measurements were made at 10-degree increments in both the common carotid artery and the superior mesenteric artery in normal subjects. Peak systolic velocity in the common carotid artery varied from 86 cm/sec at 30 degrees to 168 m/sec at 80 degrees. Over the same transducer angle variation mean velocity ranged from 28 to 53 cm/sec. Similar changes in the superior mesenteric artery flow velocities were observed by varying the angle of insonation, where peak systolic velocity varied from 108 cm/sec (30 degrees) to 280 cm/sec (80 degrees), and mean velocity ranged from 29 cm/sec (30 degrees) to 71 cm/sec (80 degrees). Measurements taken from 70 to 80 degrees produced the most dramatic deviation from those taken at 60 degrees. In the common carotid artery the 70- and 80-degree angles produced 14% and 59% increases, respectively, in peak systolic velocity and 16% and 63% increases, respectively, in mean velocity. In the superior mesenteric artery 70-degree and 80-degree angles produced 16% and 120% increases, respectively, in peak systolic velocity and 17% and 111% increases, respectively, in mean velocity. At 80 degrees the percent increases in measured flow velocities for the superior mesenteric artery were significantly greater than those for the common carotid artery.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

本研究旨在量化因脉冲多普勒超声探测角度变化而导致的颈总动脉和肠系膜上动脉双功超声动脉血流速度(厘米/秒)的变化。采用双功扫描测量30度至80度角度范围内的收缩期峰值流速和平均流速;在正常受试者的颈总动脉和肠系膜上动脉中,以10度增量进行单独测量。颈总动脉的收缩期峰值流速在30度时为86厘米/秒,在80度时为168厘米/秒。在相同的换能器角度变化范围内,平均流速为28至53厘米/秒。通过改变探测角度,观察到肠系膜上动脉血流速度有类似变化,其中收缩期峰值流速从108厘米/秒(30度)变化到280厘米/秒(80度),平均流速从29厘米/秒(30度)变化到71厘米/秒(80度)。70度至80度的测量结果与60度时的测量结果偏差最大。在颈总动脉中,70度和80度角度时收缩期峰值流速分别增加了14%和59%,平均流速分别增加了16%和63%。在肠系膜上动脉中,70度和80度角度时收缩期峰值流速分别增加了16%和120%,平均流速分别增加了17%和111%。在80度时,肠系膜上动脉测量流速的增加百分比显著大于颈总动脉。(摘要截选至250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验