Suppr超能文献

涤纶人工血管中超声血流定量测量

Quantitative ultrasonic blood flow measurement in Dacron grafts.

作者信息

LoGerfo F W, Corson J D

出版信息

Surgery. 1976 May;79(5):569-72.

PMID:131380
Abstract

A quantitative directional Doppler system was used to determine blood flow in 8 mm. aortoaortic Dacron grafts in dogs. The Doppler was calibrated with frequency standards and the output was integrated for one minute intervals. With the probe held at an angle of 60 degrees to the flow axis, the average frequency of the Doppler shift was determined. Frequency was converted to velocity with the Doppler equation. Volume flow was calculated from the product of velocity and cross-sectional area of the graft. When compared with an electromagnetic flowmeter, the Doppler system was highly accurate with flow rates up to 900 c.c per minute. At flow rates above 900 c.c. per minute, there was a significant decrease in Doppler output which probably is due to the effects of turbulence and background noise on the Doppler signal.

摘要

使用定量定向多普勒系统测定犬8毫米主动脉-主动脉涤纶移植血管中的血流。多普勒仪用频率标准进行校准,输出结果按一分钟的间隔进行积分。将探头以与血流轴成60度的角度固定,测定多普勒频移的平均频率。利用多普勒方程将频率转换为速度。根据速度与移植血管横截面积的乘积计算体积流量。与电磁流量计相比,多普勒系统在流速高达每分钟900立方厘米时具有很高的准确性。在流速高于每分钟900立方厘米时,多普勒输出显著下降,这可能是由于湍流和背景噪声对多普勒信号的影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验