用脉冲超声多普勒仪评估主动脉冠状动脉搭桥血管的血流情况。

Estimation of flow in aortocoronary grafts with a pulsed ultrasound Doppler meter.

作者信息

Segadal L, Matre K, Engedal H, Resch F, Grip A

出版信息

Thorac Cardiovasc Surg. 1982 Oct;30(5):265-8. doi: 10.1055/s-2007-1022403.

Abstract

A newly developed pulsed ultrasound Doppler meter was used for measurement of blood flow in aortocoronary vein grafts during operation. The results were compared with measurements obtained with conventional electromagnetic flowmetry. In 27 grafts, excellent agreement was found between electromagnetic flow probes thoroughly calibrated for varying hematocrit on fresh veins in vitro, and a clip-on type of Doppler probe (r = 0.86). In vitro calibration showed a close correspondence (r = 0.98) with the Doppler technique with no dependency on hematocrit and no need for zero calibration. The use of a conventional electromagnetic flowmeter showed strong dependency on recent calibration, both for saline and for varying hematocrit. Zero-calibration was necessary for every single graft measurement. The application of ultrasound Doppler meters of high quality together with clip-on probes of proper design proved to be superior to electromagnetic flowmetry for intraoperative blood flow measurements.

摘要

一种新研发的脉冲超声多普勒仪被用于在手术过程中测量主动脉冠状动脉静脉移植物中的血流。将结果与使用传统电磁血流仪获得的测量结果进行比较。在27个移植物中,发现经过体外新鲜静脉上针对不同血细胞比容进行全面校准的电磁血流探头与夹式多普勒探头之间具有极好的一致性(r = 0.86)。体外校准显示与多普勒技术具有密切的对应关系(r = 0.98),不依赖于血细胞比容且无需零点校准。使用传统电磁血流仪显示出对近期校准的强烈依赖性,无论是对于生理盐水还是对于不同的血细胞比容。每次移植物测量都需要进行零点校准。事实证明,高质量的超声多普勒仪与设计合适的夹式探头一起应用,在术中血流测量方面优于电磁血流仪。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验