Inada T, Tsuji H, Hayakawa Y, Maruhashi A, Tsujii H
Proton Medical Research Center, University of Tsukuba.
Nihon Igaku Hoshasen Gakkai Zasshi. 1992 Aug 25;52(8):1161-7.
A sensitive strain gauge was used to detect the movement of the chest wall in order to synchronize an irradiation control system with the respiratory cycle. The output timing signal from the system is transferred to the proton accelerator for synchronized irradiation. The timing signal is set during the expiratory phase at a duration of 1.5 to 2 seconds. The efficacy of this method was evaluated by dose volume histogram based on the treatment planning program with CT images. The volume spared by this novel method was calculated in several cases, and results suggest that the method was highly effective.
使用一个灵敏的应变仪来检测胸壁的运动,以便使照射控制系统与呼吸周期同步。来自该系统的输出定时信号被传输到质子加速器以进行同步照射。定时信号在呼气阶段设置,持续时间为1.5至2秒。基于带有CT图像的治疗计划程序,通过剂量体积直方图评估了该方法的有效性。在几个病例中计算了这种新方法所保留的体积,结果表明该方法非常有效。