Suppr超能文献

[临床放射学中的新型增感屏。第一篇——实验研究(作者译)]

[New intensifying screens in clinical radiology. Paper I-Experimental investigations (author's transl)].

作者信息

Hagemann G, Töllner D, Saure D, Freyschmidt J

出版信息

Rofo. 1976 May;124(5):483-9. doi: 10.1055/s-0029-1230374.

Abstract

The degree of intensification and resolution was compared using rare-earth-screens and conventional calcium tungstate screens. -In order to approximate usual phantom (DIN 6, 815) was equipped with a lead grid at a depth of 6 cm. Nine screen-film combinations were exposed at 70 to 80 kv., using a 12-pulse generator and Bucky grid till blackening of S = 1 was obtained. -Resolution depends substantially on the inherent noise of the system and improves from 0.6 Lp/mm. for the most sensitive combinations (intensitification factor 8.0) to 1.15 Lp/mm (intensification factor 1.7). Compared with the standard combination with a resolution of about 0.9 Lp/mm., dose was reduced by more than 40% with an increase in resolution of 45% (combination 3M-Trimax A4/XD).

摘要

使用稀土增感屏和传统钨酸钙增感屏比较了强化程度和分辨率。——为了接近常规体模(DIN 6, 815),在6厘米深度处配备了铅栅。使用12脉冲发生器和滤线栅,在70至80千伏下对九种增感屏 - 胶片组合进行曝光,直至获得S = 1的黑化度。——分辨率在很大程度上取决于系统的固有噪声,对于最敏感的组合(增感因子8.0),分辨率从0.6线对/毫米提高到1.15线对/毫米(增感因子1.7)。与分辨率约为0.9线对/毫米的标准组合相比,剂量降低了40%以上,分辨率提高了45%(3M-Trimax A4/XD组合)。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验