Mirhabibollahi B, Brooks J L, Kroll R G
Department of Microbiology, AFRC Institute of Food Research, Shinfield, Reading, UK.
Lett Appl Microbiol. 1990 Sep;11(3):119-22. doi: 10.1111/j.1472-765x.1990.tb00138.x.
An amperometric immunosensor specific to the protein A of Staphylococcus aureus, was developed using the direct electrochemical detection of phenol produced by alkaline phosphatase from phenylphosphate. The immunosensor could detect protein A at 0.01 ng/ml and could reliably detect and quantify pure cultures of protein A-bearing Staph. aureus above 10(3) cfu/ml. A similar sensitivity of detection was obtained with samples of beef and milk.