Broderick N J, Long B, Dreesen R G, Cohen M D, Cory D A, Katz B P, Kalasinski L A
Department of Radiology, Riley Hospital for Children, Indiana University School of Medicine, Indianapolis.
Pediatr Radiol. 1992;22(5):346-9. doi: 10.1007/BF02016253.
A controlled prospective study, in an animal model chosen to simulate portable neonatal radiography, was performed to compare the response of the Philips Computed Radiography (CR) system and conventional 200 speed film-screen (FS) to variation in imaging kVp. Acceptable images were obtained on the CR system over a very wide kVp range. In contrast the FS system produced acceptable images over a narrow kVp range. This ability suggests that the CR system should eliminate the need for repeat examinations in cases where a suboptimal kVp setting would have resulted in an unacceptable FS image. CR technology should therefore be ideally suited to portable radiography especially in situations where selection of correct exposure factors is difficult as in the neonatal nursery.
为了比较飞利浦计算机放射成像(CR)系统和传统的200速屏-片(FS)对成像千伏峰值(kVp)变化的响应,我们在一个被选来模拟便携式新生儿放射成像的动物模型上进行了一项对照前瞻性研究。在很宽的kVp范围内,CR系统都能获得可接受的图像。相比之下,FS系统只能在较窄的kVp范围内产生可接受的图像。这种能力表明,在次优kVp设置会导致FS图像不可接受的情况下,CR系统应该可以消除重复检查的必要性。因此,CR技术应该非常适合便携式放射成像,特别是在新生儿重症监护室等难以选择正确曝光参数的情况下。