采用双图像记录技术对传统放射摄影和计算机放射摄影进行直接比较。

Direct comparison of conventional and computed radiography with a dual-image recording technique.

作者信息

MacMahon H, Sanada S, Doi K, Giger M, Xu X W, Yin F F, Montner S M, Carlin M

机构信息

Kurt Rossmann Laboratories for Radiologic Image Research, Department of Radiology, University of Chicago, IL 60637.

出版信息

Radiographics. 1991 Mar;11(2):259-68. doi: 10.1148/radiographics.11.2.2028063.

Abstract

To compare the image quality of computed radiographic (CR) with conventional screen-film images, the authors used a dual-image recording technique. Images were simultaneously acquired with a conventional screen-film combination and a storage-phosphor imaging plate loaded into a single cassette. Wiener spectra and modulation transfer function were compared for both image types. A preliminary observer performance test was conducted with chest images obtained with the dual-image recording technique on portable and fixed equipment. Analysis of physical parameters and observer test results suggests that the conventional screen-film system can provide slightly superior image quality, although the CR system has some advantage in bedside applications.

摘要

为比较计算机X线摄影(CR)与传统屏-片图像的图像质量,作者采用了双图像记录技术。图像通过传统屏-片组合和装入单个暗盒的存储磷光成像板同时采集。比较了两种图像类型的维纳频谱和调制传递函数。利用双图像记录技术在便携式和固定式设备上获得胸部图像,进行了初步的观察者性能测试。物理参数分析和观察者测试结果表明,传统屏-片系统可提供略优的图像质量,尽管CR系统在床边应用方面具有一定优势。