Direct comparison of conventional and computed radiography with a dual-image recording technique.

作者信息

MacMahon H, Sanada S, Doi K, Giger M, Xu X W, Yin F F, Montner S M, Carlin M

机构信息

Kurt Rossmann Laboratories for Radiologic Image Research, Department of Radiology, University of Chicago, IL 60637.

出版信息

Radiographics. 1991 Mar;11(2):259-68. doi: 10.1148/radiographics.11.2.2028063.

Abstract

To compare the image quality of computed radiographic (CR) with conventional screen-film images, the authors used a dual-image recording technique. Images were simultaneously acquired with a conventional screen-film combination and a storage-phosphor imaging plate loaded into a single cassette. Wiener spectra and modulation transfer function were compared for both image types. A preliminary observer performance test was conducted with chest images obtained with the dual-image recording technique on portable and fixed equipment. Analysis of physical parameters and observer test results suggests that the conventional screen-film system can provide slightly superior image quality, although the CR system has some advantage in bedside applications.