Animal implantation results with the Utah-100 total artificial heart.

作者信息

Kinoshita M, Pantalos G M, Holfert J W, Dew P A, Burns G L, Mohammad S F, Olsen D B

机构信息

Artificial Heart Research Laboratory, University of Utah, Salt Lake City 84103-1414.

出版信息

ASAIO J. 1992 Apr-Jun;38(2):108-12.

PMID:1421602
Abstract

The Utah-100 total artificial heart was designed to have increased reliability over the Jarvik-7 total artificial heart, achieve better fit, and minimize device associated thrombus formation, without decreasing the function. The Utah-100 heart was tested in 28 calves and 3 sheep. The smallest animal at the time of implantation weighted 54 kg. Mean survival duration was 78 days (range, 1-331 days), with 14 animals surviving longer than 60 days. Multiple organ function was maintained satisfactorily with the Utah-100 artificial heart, and mean plasma free hemoglobin values in the calves that survived longer than 100 days were less than 10 mg/dl. Hemorrhage was the main cause of death in animals dying within 30 days after implantation (5/13, 38%); infection was another primary cause of death or termination (4/31, 13%). Deaths due to mechanical failure occurred from valve or diaphragm failure in two cases, yielding a 91% reliability at a 90% confidence level for 60 days' support. No animal died because of driver or other technical failure. Utah-100 hearts showed superior antithrombogenicity in the connector and valve-related areas when compared with the results of the Jarvik-7 heart, which was also fabricated and implanted in our laboratory (p less than 0.01). With these test results, the authors anticipate that the Utah-100 heart will be a safe and effective device for interim use as a bridge to heart transplantation.

摘要

相似文献

1
Animal implantation results with the Utah-100 total artificial heart.
ASAIO J. 1992 Apr-Jun;38(2):108-12.
2
Design mediated thrombus reduction in the Utah-100 total artificial heart.
ASAIO Trans. 1990 Jul-Sep;36(3):M230-4.
3
CardioWest (Jarvik) total artificial heart: a single-center experience with 42 patients.CardioWest(贾维克)全人工心脏:42例患者的单中心经验。
Ann Thorac Surg. 2009 Jan;87(1):124-9; discussion 130. doi: 10.1016/j.athoracsur.2008.09.048.
4
Temporary use of the Jarvik-7 total artificial heart before transplantation.
N Engl J Med. 1987 Jan 15;316(3):130-4. doi: 10.1056/NEJM198701153160303.
5
Summary of the clinical use of the Symbion total artificial heart: a registry report.共生全人工心脏临床应用总结:一项注册报告
J Heart Lung Transplant. 1992 Jan-Feb;11(1 Pt 1):103-16.
6
Mechanical failures of the pneumatic Utah-100 and Jarvik total artificial hearts. A comparative study.犹他100型气动式和贾维克全人工心脏的机械故障。一项对比研究。
ASAIO Trans. 1989 Jul-Sep;35(3):697-9. doi: 10.1097/00002480-198907000-00171.
7
Development and evaluation of a swine model to assess the preclinical safety of mechanical heart valves.用于评估机械心脏瓣膜临床前安全性的猪模型的开发与评估。
J Heart Valve Dis. 2000 Sep;9(5):710-9; discussion 719-20.
8
Bridge to transplantation with the CardioWest total artificial heart: the international experience 1993 to 1995.使用CardioWest全人工心脏过渡到移植:1993年至1995年的国际经验
J Heart Lung Transplant. 1996 Jan;15(1 Pt 1):94-9.
9
The initial surgical intensive care unit nursing experience with the permanent total artificial heart.
Heart Lung. 1986 Sep;15(5):475-87.
10
Calcification of the driving diaphragm in a total artificial heart.全人工心脏中驱动隔膜的钙化
Czech Med. 1987;10(1):16-28.