正电子发射断层扫描可检测心肌中具有持续24小时单光子发射计算机断层扫描201Tl缺损的代谢活性。

Positron emission tomography detects metabolic viability in myocardium with persistent 24-hour single-photon emission computed tomography 201Tl defects.

作者信息

Brunken R C, Mody F V, Hawkins R A, Nienaber C, Phelps M E, Schelbert H R

机构信息

Department of Radiological Sciences, UCLA School of Medicine 90024-1721.

出版信息

Circulation. 1992 Nov;86(5):1357-69. doi: 10.1161/01.cir.86.5.1357.

Abstract

BACKGROUND

Four-hour 201Tl redistribution images underestimate myocardial viability in patients with coronary artery disease (CAD). Because 4-hour defects often redistribute late, delayed imaging may enhance assessment of tissue viability. Myocardial metabolic activity was therefore assessed with positron emission tomography (PET) in 26 CAD patients with impaired ventricular function (ejection fraction, 32.1 +/- 13.9%) and 24-hour single-photon emission computed tomography (SPECT) 201Tl defects.

METHODS AND RESULTS

On circumferential profile analysis, PET ischemia was defined by preserved glucose metabolism in hypoperfused myocardium, and PET infarction was defined by concordant reductions in perfusion and metabolism. On 19 stress-redistribution and seven rest-redistribution SPECT studies, four observers visually scored 201Tl activity in eight segments on a scale from 0 (normal) to 3 (complete defect). Using an improvement in visual score > or = 0.75 to define redistribution, there were 100 fixed, 17 partially reversible, and 12 completely reversible defects. PET identified tissue metabolic activity in 51 (51%) segments with fixed defects (21 PET ischemia, 30 PET normal) and nine (53%) segments with partially reversible defects (five PET ischemia, four PET normal). When grouped by 24-hour score, the proportion of fixed defects with metabolic activity varied from 84% (scores < or = 1.4) to 15% (scores > 2.6). For partially reversible defects, only 53% with scores < 2.0 and one of two with scores > or = 2.0 were considered metabolically viable on PET. Of 12 completely reversible defects, six (50%) were normal, five (42%) had PET ischemia, and one (8%) had PET infarction. The proportion of fixed defects with metabolic activity did not depend on whether a rest or stress study was performed or on the change in visual score used to define 201Tl redistribution (0.25, 0.50, 0.75, and 1.00).

CONCLUSIONS

In CAD patients, PET identifies glucose metabolic activity in the majority of fixed 24-hour 201Tl defects. However, very severe (near-complete) 24-hour 201Tl defects are less likely to exhibit metabolic activity on PET imaging than are defects with less-pronounced reductions in segmental 201Tl activity.

摘要

背景

对于冠心病(CAD)患者,4小时铊-201(201Tl)再分布影像会低估心肌存活性。由于4小时的缺损常常在后期再分布,延迟显像可能会增强对组织存活性的评估。因此,对26例心室功能受损(射血分数为32.1±13.9%)且有24小时单光子发射计算机断层扫描(SPECT)201Tl缺损的CAD患者,采用正电子发射断层扫描(PET)评估心肌代谢活性。

方法与结果

在圆周轮廓分析中,PET缺血定义为灌注减低的心肌中葡萄糖代谢保留,PET梗死定义为灌注和代谢均一致降低。在19例负荷-再分布和7例静息-再分布SPECT研究中,4名观察者对8个节段的201Tl活性进行视觉评分,范围从0(正常)至3(完全缺损)。使用视觉评分改善≥0.75来定义再分布,有100个固定缺损、17个部分可逆缺损和12个完全可逆缺损。PET在51个(51%)有固定缺损的节段(21个PET缺血,30个PET正常)和9个(53%)有部分可逆缺损的节段(5个PET缺血,4个PET正常)中识别出组织代谢活性。按24小时评分分组时,有代谢活性的固定缺损比例从84%(评分≤1.4)至15%(评分>2.6)不等。对于部分可逆缺损,PET上评分<2.0的只有53%以及评分≥2.0的两个中的一个被认为有代谢活性。在12个完全可逆缺损中,6个(50%)正常,5个(42%)有PET缺血,1个(8%)有PET梗死。有代谢活性的固定缺损比例不取决于进行的是静息还是负荷研究,也不取决于用于定义201Tl再分布的视觉评分变化(0.25、0.50、0.75和1.00)。

结论

在CAD患者中,PET在大多数24小时固定的201Tl缺损中识别出葡萄糖代谢活性。然而,与节段性201Tl活性降低不那么明显的缺损相比,非常严重(近乎完全)的24小时201Tl缺损在PET显像上显示代谢活性的可能性较小。