使用传统聚合物除颤垫对人类志愿者进行经皮低阻抗心脏起搏的有效性和安全性。

Efficacy and safety of transcutaneous low-impedance cardiac pacing in human volunteers using conventional polymeric defibrillation pads.

作者信息

Chapman P D, Stratbucker R A, Schlageter D P, Pruzina S P

机构信息

Department of Cardiology, Medical College of Wisconsin, Milwaukee.

出版信息

Ann Emerg Med. 1992 Dec;21(12):1451-3. doi: 10.1016/s0196-0644(05)80058-7.

Abstract

STUDY OBJECTIVES

To assess the safety and efficacy of transcutaneous cardiac pacing using low-impedance defibrillation-type, self-adhesive polymer electrode pads positioned in the same anatomic sites typical of such pad placement in emergency defibrillation attempts.

DESIGN

Prospective, randomized, single-blinded normal subject investigation.

METHODS

Thirty healthy unmedicated adult volunteers of both sexes were paced transcutaneously to the threshold of capture and beyond by an intensity factor of 125%. A commercial defibrillator-pacer operating in demand mode had its rate set-point higher than the subjects' base-line rate by 125%. Threshold was established as the pulse current required to effect 75% captured beats. Pacing was continued at the 125% level for two minutes, during which subjective discomfort levels were recorded on a 1-to-5 pain scale.

RESULTS

All subjects but one were able to complete the study as designed. The single exception complained of intolerable pain early in the two-minute trial. Capture current had a mean value of 80 mA while the pain assessment averaged 3.2 on our five-point scale. No subject showed any untoward cardiovascular effects either during or after the study.

CONCLUSION

In normal human volunteers, transcutaneous cardiac pacing with low-impedance polymer pad electrodes is safe. By extrapolation from this study population, it is sufficiently effective to accommodate successfully the vast majority of clinical circumstances likely to be encountered in emergency medicine.

摘要

研究目的

评估使用低阻抗除颤型自粘性聚合物电极片进行经皮心脏起搏的安全性和有效性,这些电极片放置在紧急除颤尝试中此类电极片典型放置的相同解剖部位。

设计

前瞻性、随机、单盲正常受试者研究。

方法

30名未用药的健康成年男女志愿者接受经皮起搏,直至捕获阈值,并超过该阈值125%的强度因子。一台按需模式运行的商用除颤起搏器的心率设定点比受试者的基线心率高125%。阈值确定为实现75%捕获搏动所需的脉冲电流。在125%的水平上继续起搏两分钟,在此期间,主观不适程度按照1至5级疼痛量表进行记录。

结果

除一名受试者外,所有受试者均能按设计完成研究。唯一的例外是在两分钟试验早期抱怨疼痛难忍。捕获电流的平均值为80毫安,而我们五点量表上的疼痛评估平均为3.2。在研究期间或之后,没有受试者出现任何不良心血管效应。

结论

在正常人类志愿者中,使用低阻抗聚合物垫电极进行经皮心脏起搏是安全的。从该研究人群推断,它足够有效,能够成功适应急诊医学中可能遇到的绝大多数临床情况。