Bohn W W
University of Missouri, Research Medical Center, Kansas City 64132.
Clin Orthop Relat Res. 1992 Dec(285):155-7.
Removal of a modular head femoral prosthesis may be extremely difficult during revision total hip arthroplasty (THA). A universal modular stem extractor was developed that achieves a secure attachment to the stem taper, and applies tensile forces with a five-pound slaphammer. Clinical use indicates that this new device facilitates uncomplicated modular femoral stem removal.
在翻修全髋关节置换术(THA)过程中,取出模块化股骨头假体可能极其困难。研发了一种通用的模块化柄取出器,它能牢固地附着在柄的锥度部位,并通过一个五磅重的冲击锤施加拉力。临床应用表明,这种新装置有助于顺利取出模块化股骨干。