Suppr超能文献

双谱指数、基于外源性输入的听觉诱发电位自回归模型以及预测的丙泊酚浓度在丙泊酚与瑞芬太尼麻醉期间测量患者反应性的能力。

Ability of the bispectral index, autoregressive modelling with exogenous input-derived auditory evoked potentials, and predicted propofol concentrations to measure patient responsiveness during anesthesia with propofol and remifentanil.

作者信息

Struys Michel M R F, Vereecke Hugo, Moerman Annelies, Jensen Erik Weber, Verhaeghen David, De Neve Nicolaas, Dumortier Frank J E, Mortier Eric P

机构信息

Department of Anesthesia, Ghent University Hospital, Belgium.

出版信息

Anesthesiology. 2003 Oct;99(4):802-12. doi: 10.1097/00000542-200310000-00010.

Abstract

BACKGROUND

This study was conducted to compare the performance accuracy of the independent variables Bispectral Index (BIS), A-Line ARX index (AAI), and predicted propofol effect-site concentration (CePROP) to measure the dependent variables of loss of responses to different stimulation defined as loss of response to verbal command (LORverbal), eyelash reflex (LORlash), and noxious stimulus (LORnoxious) during stepwise increased levels of propofol infusion with and without remifentanil.

METHODS

Forty-five patients were randomly allocated to one of three groups (0, 2, and 4 ng/ml remifentanil) to receive graded CePROP and predicted effect compartment controlled remifentanil (CeREMI). At every step, the ability to respond to verbal command using the Observer's Assessment of Alertness/Sedation Scale (OAA/S), eyelash reflex, and electrical tetanic noxious stimulus were compared against BIS, AAI, and CePROP. Prediction probability and sensitivity/specificity were calculated.

RESULTS

Increasing CeREMI increased BIS and AAI values at LORverbal and LORlash and decreased CePROP. Similar findings were found for LORnoxious. The overall prediction probability to measure the hypnotic component of anesthesia remained accurate in the three groups for BIS, AAI, and CePROP. Combined information from CePROP, CeREMI, and BIS or AAI increased the overall prediction probability for predicting the OAA/S scale and LORlash. Less accuracy to LORnoxious was found in all independent variables.

CONCLUSIONS

Although BIS, AAI, and CePROP were influenced by remifentanil during propofol administration, their ability to detect OAA/S and LORlash remained accurate. Improved performance is obtained when BIS and AAI are measured in conjunction with drug targeted effect-site concentrations. Remifentanil decreases the ability of these independent variables to detect LORnoxious.

摘要

背景

本研究旨在比较双谱指数(BIS)、A线ARX指数(AAI)和预测丙泊酚效应室浓度(CePROP)这些自变量在测量不同刺激反应消失(定义为对语言指令反应消失[LORverbal]、睫毛反射消失[LORlash]和有害刺激反应消失[LORnoxious])时的性能准确性,这些刺激反应消失是在有和没有瑞芬太尼的情况下逐步增加丙泊酚输注水平时出现的。

方法

45例患者被随机分配到三组(0、2和4 ng/ml瑞芬太尼)之一,接受分级的CePROP和预测效应室控制的瑞芬太尼(CeREMI)。在每一步,使用观察者警觉/镇静评分量表(OAA/S)对语言指令的反应能力、睫毛反射和电强直有害刺激与BIS、AAI和CePROP进行比较。计算预测概率和敏感性/特异性。

结果

增加CeREMI会使LORverbal和LORlash时的BIS和AAI值增加,CePROP降低。LORnoxious时也有类似发现。在三组中,BIS、AAI和CePROP测量麻醉催眠成分的总体预测概率仍然准确。来自CePROP、CeREMI以及BIS或AAI的综合信息增加了预测OAA/S量表和LORlash的总体预测概率。在所有自变量中,对LORnoxious的准确性较低。

结论

尽管在丙泊酚给药期间BIS、AAI和CePROP受瑞芬太尼影响,但它们检测OAA/S和LORlash的能力仍然准确。当BIS和AAI与药物靶向效应室浓度一起测量时,性能得到改善。瑞芬太尼降低了这些自变量检测LORnoxious的能力。