Hillman Todd A, Shelton Clough
Division of Otolaryngology, University of Utah, 50 North Medical Drive, Salt Lake City, UT 84132, USA.
Laryngoscope. 2003 Oct;113(10):1731-5. doi: 10.1097/00005537-200310000-00013.
To compare the complication rate and hearing results of a new, lightweight, titanium ossicular replacement prosthesis with Plastipore prostheses (Xomed, Jacksonville, FL).
Retrospective.
Charts were reviewed for type of operation, type of prosthesis used, extrusion rate, prostheses failure rate, and hearing thresholds at multiple frequencies and at multiple follow-up points. The dependant variable for hearing results was the four-frequency average air-bone gap.
There were 84 patients undergoing tympanoplasty with the Plastipore prosthesis and 53 with the titanium. There was one extrusion in the titanium group. There was an additional single incidence of prosthesis failure in the titanium group. Overall hearing results were comparable with an air-bone gap average of 19.3 dB in the Plastipore group compared with the titanium group with an air-bone gap of 22.0 dB (P =.08). Sixty percent of patients had a postoperative air-bone gap of 20 dB or less in the Plastipore group. In the titanium group, 45.3% achieved a 20 dB or less postoperative air-bone gap. Plastipore had a lower air-bone gap than the titanium when a canal wall up operation was performed (17.8 vs. 23.9 dB) and tended toward a lower air-bone gap when a total ossicular prosthesis was needed (22 vs. 27 dB) (P <.07).
The titanium prosthesis is a new ossicular replacement prosthesis that provides excellent visualization during insertion and provides hearing results that are comparable with Plastipore. The Plastipore prosthesis performed better in canal wall up mastoidectomy situations and tended toward better performance when a total ossicular replacement prosthesis was needed.
比较一种新型轻质钛质听骨置换假体与Plastipore假体(Xomed公司,佛罗里达州杰克逊维尔)的并发症发生率及听力结果。
回顾性研究。
查阅病历,了解手术类型、所用假体类型、脱出率、假体故障率以及多个频率和多个随访点的听力阈值。听力结果的因变量为四频率平均气骨导差。
84例患者接受了使用Plastipore假体的鼓室成形术,53例使用钛质假体。钛质假体组有1例脱出。钛质假体组还有1例假体故障。总体听力结果相当,Plastipore组平均气骨导差为19.3dB,钛质假体组为22.0dB(P = 0.08)。Plastipore组60%的患者术后气骨导差为20dB或更小。在钛质假体组,45.3%的患者术后气骨导差达到20dB或更小。行外耳道上壁手术时,Plastipore的气骨导差低于钛质假体(17.8对23.9dB),在需要全听骨假体时,Plastipore的气骨导差也有降低趋势(22对27dB)(P < 0.07)。
钛质假体是一种新型听骨置换假体,植入时视野良好,听力结果与Plastipore相当。在耳道上壁乳突切除术中,Plastipore假体表现更佳,在需要全听骨置换假体时,其性能也有更好的趋势。