van Holten Jacqueline, Wielopolski Piotr, Brück Ekkes, Pattynama Peter M T, van Dijk Lukas C
Department of Radiology, University Medical Center, Rotterdam, The Netherlands.
Magn Reson Med. 2003 Oct;50(4):879-83. doi: 10.1002/mrm.10600.
Intraluminal stent signal characterization by MRI is generally hampered by signal loss from the metallic stent material. This signal loss is related to magnetic susceptibility and RF shielding. Even when stent materials with low magnetic susceptibility are used, RF shielding can still be problematic. In this article we have shown that high flip angle imaging enables morphology assessment and tissue characterization in stents made of stainless steel 316L, NiTinol, and ABI-alloy.