Suppr超能文献

比利时使用FilterWire EX预防颈动脉支架置入术中栓塞事件的经验。

Belgian experience with FilterWire EX in the prevention of embolic events during carotid stenting.

作者信息

Bosiers Marc, Peeters Patrick, Verbist Jurgen, Schroë Herman, Deloose Koen, Lauwers Geert, Stockx Luc

机构信息

Department of Vascular Surgery, AZ St-Blasius, Dendermonde.

出版信息

J Endovasc Ther. 2003 Aug;10(4):695-701. doi: 10.1177/152660280301000403.

Abstract

PURPOSE

To evaluate the performance of the FilterWire EX for the capture and removal of debris and prevention of embolic events during endovascular carotid interventions.

METHODS

In the period from October 2001 to May 2002, 100 patients (60 men; mean age 72 years, range 48-90), 69 symptomatic with >or=50% internal carotid artery (ICA) stenosis and 31 asymptomatic with >or=70% stenosis were scheduled for carotid stenting with cerebral protection provided by the FilterWire EX system. Satisfactory performance of the filter was defined as effectiveness in preventing intraprocedural embolic events shown by visible debris in the filter and unchanged neurological status assessed by an independent neurologist.

RESULTS

The placement and retrieval of the FilterWire EX was technically successful in 93 of 100 patients. Five delivery failures were due to severe ICA angulation; one stenosis was underestimated on preprocedural imaging, and one case was aborted following a major stroke during sheath manipulation before filter introduction. In 56.9% of the protected cases, the filter contained debris. One neurological event occurred 6 hours after the procedure: a transient ischemic attack from which the patient recovered fully after 2 hours. No other procedure-related events were recorded within 30 days of the procedure. The procedural success among the 93 patients stented under FilterWire EX protection was 98% (92% by intention to treat).

CONCLUSION

The FilterWire EX appears to be an effective tool in the prevention of embolic events and in preserving neurological status during endovascular treatment of carotid artery stenosis.

摘要

目的

评估FilterWire EX在血管内颈动脉介入治疗期间捕获和清除碎片以及预防栓塞事件的性能。

方法

在2001年10月至2002年5月期间,100例患者(60例男性;平均年龄72岁,范围48 - 90岁),69例有症状,颈内动脉(ICA)狭窄≥50%,31例无症状,狭窄≥70%,计划采用FilterWire EX系统提供脑保护的情况下进行颈动脉支架置入术。滤网的满意性能定义为在预防术中栓塞事件方面有效,表现为滤网上可见碎片,且经独立神经科医生评估神经状态无变化。

结果

FilterWire EX的放置和取出在100例患者中的93例技术上成功。5次输送失败是由于ICA严重成角;1例术前成像低估了狭窄程度,1例在滤网引入前鞘管操作期间发生重大中风后手术中止。在56.9%的受保护病例中,滤网含有碎片。术后6小时发生1例神经事件:短暂性脑缺血发作,患者在2小时后完全康复。在术后30天内未记录到其他与手术相关的事件。在FilterWire EX保护下进行支架置入的93例患者中,手术成功率为98%(意向性治疗为92%)。

结论

FilterWire EX似乎是预防颈动脉狭窄血管内治疗期间栓塞事件和保持神经状态的有效工具。