Early failure mechanisms of constrained tripolar acetabular sockets used in revision total hip arthroplasty.

作者信息

Cooke Christopher C, Hozack William, Lavernia Carlos, Sharkey Peter, Shastri Shani, Rothman Richard H

机构信息

Orthopaedic Surgery, Lancaster, Pennsylvania, USA.

出版信息

J Arthroplasty. 2003 Oct;18(7):827-33. doi: 10.1016/s0883-5403(03)00325-5.

Abstract

Fifty-eight patients received an Osteonics constrained acetabular implant for recurrent instability (46), girdlestone reimplant (8), correction of leg lengthening (3), and periprosthetic fracture (1). The constrained liner was inserted into a cementless shell (49), cemented into a pre-existing cementless shell (6), cemented into a cage (2), and cemented directly into the acetabular bone (1). Eight patients (13.8%) required reoperation for failure of the constrained implant. Type I failure (bone-prosthesis interface) occurred in 3 cases. Two cementless shells became loose, and in 1 patient, the constrained liner was cemented into an acetabular cage, which then failed by pivoting laterally about the superior fixation screws. Type II failure (liner locking mechanism) occurred in 2 cases. Type III failure (femoral head locking mechanism) occurred in 3 patients. Seven of the 8 failures occurred in patients with recurrent instability. Constrained liners are an effective method for treatment during revision total hip arthroplasty but should be used in select cases only.