Jones Christopher W, Malahias Michael-Alexander, Baral Elexis, Wright Timothy, Sculco Thomas P, Sculco Peter K
Hospital for Special Surgery, New York, NY, USA.
National and Kapodistrian University of Athens, School of Medicine, Athens, Greece.
Arthroplast Today. 2018 May 18;4(3):270-274. doi: 10.1016/j.artd.2018.04.002. eCollection 2018 Sep.
Constrained acetabular liners have been developed for patients who are at high risk for dislocation or who are undergoing revision surgery for recurrent dislocations. We report on 2 cases of failure of tripolar constrained liners due to severe backside polyethylene wear after dissociation of the outer polyethylene liner without dislocation, a mode of failure not previously reported. The backside of the inserts suffered severe polyethylene deformation, wear, and scratching due to dissociation from the locking mechanism. In patients with tripolar constrained liners, radiographic evidence of eccentric wear should be considered as possible occult dissociation of the polyethylene liner within the shell. Conversion to a modular dual mobility liner appears to be a viable solution in this setting.
对于脱位风险高或因复发性脱位而接受翻修手术的患者,已开发出受限髋臼衬垫。我们报告了2例三极受限衬垫失败的病例,原因是外部聚乙烯衬垫解离后出现严重的后侧聚乙烯磨损,且未发生脱位,这种失败模式此前未见报道。由于与锁定机制解离,插入物的后侧出现严重的聚乙烯变形、磨损和刮擦。对于使用三极受限衬垫的患者,偏心磨损的影像学证据应被视为聚乙烯衬垫在髋臼壳内可能发生隐匿性解离。在这种情况下,转换为模块化双动衬垫似乎是一种可行的解决方案。