Physical characteristics of low and medium energy collimators for 123I imaging and simultaneous dual-isotope imaging.

作者信息

Inoue Y, Shirouzu I, Machida T, Yoshizawa Y, Akita F, Doi I, Watadani T, Noda M, Yoshikawa K, Ohtomo K

机构信息

Department of Radiology, Institute of Medical Science, University of Tokyo, Japan.

出版信息

Nucl Med Commun. 2003 Nov;24(11):1195-202. doi: 10.1097/00006231-200311000-00011.

Abstract

Septal penetration of high energy photons may degrade 123I images obtained with a low energy collimator. We evaluated the physical characteristics of a low energy, high resolution (LEHR) collimator, special LEHR (SLEHR) collimator, and medium energy collimator for 123I studies. The cross-talk of 99mTc and 123I into the 201Tl window was also examined. Sensitivity and spatial resolution were measured with each collimator. Point sources of 99mTc and 123I were imaged at various source-collimator distances using multiple energy windows, and the effects of collimator choice on energy spectrum and spatial distribution of photopeak counts were assessed. For 99mTc, both sensitivity and resolution were similar with the low energy collimators, and higher sensitivity and lower resolution were observed with the medium energy collimator. For 123I, the full width at tenth maximum was larger for 123I than for 99mTc when using the LEHR collimator. Acquisitions with multiple energy windows revealed severe degradation due to septal penetration in imaging 123I with the LEHR collimator, especially at short distances. The degradation was reduced with the SLEHR collimator and further with the medium energy collimator. In both 99mTc and 123I imaging, cross-talk into the 201Tl window was larger at shorter distances and the largest with the LEHR collimator. In conclusion, variation in collimator geometry causes differences in the effect of septal penetration on 123I images and in cross-talk into the 201Tl window. The SLEHR collimator may be suitable for use in high resolution 123I imaging and simultaneous 99mTc/201Tl imaging. Use of the medium energy collimator appears to be preferable in quantitative 123I studies.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验