Inoue Yusuke, Shirouzu Ichiro, Machida Toru, Yoshizawa Yasunori, Akita Fumihide, Minami Manabu, Ohtomo Kuni
Department of Radiology, Institute of Medical Science, University of Tokyo, Japan.
J Nucl Cardiol. 2004 Jul-Aug;11(4):433-9. doi: 10.1016/j.nuclcard.2004.04.009.
Septal penetration of high-energy photons may degrade the quality of single photon emission computed tomography (SPECT) of the heart with iodine 123-labeled tracers. We investigated the impact of collimator choice on cardiac SPECT with I-123.
SPECT of a thoracic phantom containing I-123 solution was performed with a low-energy high-resolution (LEHR) collimator, special LEHR (SLEHR) collimator, and medium-energy (ME) collimator, and the cavity-to-myocardium contrast, wall thickness, and defect contrast were compared among the collimators. For all indices, use of the SLEHR collimator yielded the best results. Comparison between the LEHR and ME collimators revealed that the cavity-to-myocardium contrast and contrast for large defects were better with the ME collimator, whereas wall thickness and contrast for small defects were similar. Scatter correction by the triple-energy window method improved the indices examined; however, the superiority of the SLEHR collimator was still observed after correction.
Collimator choice substantially influences the quality of cardiac SPECT with I-123-labeled agents, and an appropriate collimator needs to be selected in consideration of septal penetration and spatial resolution.
高能光子的隔肌穿透可能会降低使用碘123标记示踪剂的心脏单光子发射计算机断层扫描(SPECT)的质量。我们研究了准直器选择对I-123心脏SPECT的影响。
使用低能高分辨率(LEHR)准直器、特殊LEHR(SLEHR)准直器和中能(ME)准直器对含有I-123溶液的胸部模型进行SPECT检查,并比较各准直器之间的腔室与心肌对比度、壁厚度和缺损对比度。对于所有指标,使用SLEHR准直器产生的结果最佳。LEHR和ME准直器之间的比较显示,ME准直器的腔室与心肌对比度和大缺损对比度更好,而壁厚度和小缺损对比度相似。通过三能量窗法进行散射校正改善了所检查的指标;然而,校正后仍观察到SLEHR准直器的优势。
准直器的选择对使用I-123标记剂的心脏SPECT质量有重大影响,需要考虑隔肌穿透和空间分辨率来选择合适的准直器。