Health Phys. 2003 Nov;85(5 Suppl):S81-4. doi: 10.1097/00004032-200311002-00008.
The root cause of many false continuous air monitor (CAM) alarms is revealed for CAMs that use constant spectral shape assumptions in transuranic (TRU) alpha activity determination algorithms. This paper shows that when atmospheric radon levels continually decrease and bottom out at a minimum level, reduced false TRU count rates are not only expected but measured. Similarly, when the radon levels continually increase to a maximum level, elevated false TRU count rates were measured as predicted. The basis for expecting this dependence on changes in radon levels is discussed.