Suppr超能文献

采用石墨炉塞曼原子吸收光谱法直接测定血清中的锌。

Direct determination of zinc in serum by Zeeman atomic absorption spectrometry with a graphite furnace.

作者信息

D'Haese P C, Lamberts L V, Vanheule A O, De Broe M E

机构信息

Department of Nephrology-Hypertension, University Hospital Antwerp, University of Antwerp, Edegem, Belgium.

出版信息

Clin Chem. 1992 Dec;38(12):2439-43.

PMID:1458581
Abstract

We developed a precise and accurate graphite furnace atomic absorption spectrometric method for the direct determination of zinc in serum. Serum samples are analyzed after 20-fold dilution with water of ultrapure analytical grade. No other reagent is used, from the moment of sampling until measurement. During atomization, the argon flow is kept at 150 mL/min instead of gas stop, to decrease the sensitivity and thus allow lower dilution ratios. Zinc concentrations are determined against a serum-matched calibration curve. Graphite tubes are uncoated and no L'vov platform is used. Between-run CVs were 5.9%, 3.5%, and 1.9% for serum zinc concentrations of 0.93, 1.15, and 1.43 mg/L, respectively. The characteristic mass was 9 pg, and the detection limit (meanblank + 3SDblank) was 0.060 mg/L.

摘要

我们开发了一种精确且准确的石墨炉原子吸收光谱法,用于直接测定血清中的锌。血清样品用超纯分析级水进行20倍稀释后进行分析。从采样到测量期间,不使用其他试剂。在原子化过程中,氩气流速保持在150 mL/min而不是停止通气,以降低灵敏度,从而允许更低的稀释倍数。锌浓度通过与血清匹配的校准曲线来测定。石墨管未涂层,且未使用热解涂层石墨管平台。对于血清锌浓度分别为0.93、1.15和1.43 mg/L的情况,批间变异系数分别为5.9%、3.5%和1.9%。特征质量为9 pg,检测限(平均空白值 + 3倍空白值标准差)为0.060 mg/L。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验