Dejardin L M, Kahanovitz N, Arnoczky S P, Simon B J
Laboratory for Comparative Orthopaedic Research, College of Veterinary Medicine, Michigan State University, East Lansing, MI 48824-1314, USA.
Spine J. 2001 Sep-Oct;1(5):341-7. doi: 10.1016/s1529-9430(01)00100-0.
The use of electrical stimulation has been shown to increase the rate of successful spinal fusions. It is possible that increasing the current density of these stimulators may increase the speed and success rate of these fusions.
This study evaluated the effects of varied current densities on the speed and success rate of spinal fusion in dogs.
STUDY DESIGN/SETTING: Three different current densities (0.83 microA/cm, 4 microA/cm and 10 microA/cm) were used to stimulate spinal fusions in a canine model over a 12-week period.
Radiographic and histologic assessments were used to determine the degree of facet fusions at each time period.
Fifteen dogs underwent spinal facet fusion bilaterally at the level of L1-2 and L4-5. Each fusion site was electrically stimulated using one of three current densities. At 6, 9 and 12 weeks, the specimens were evaluated using high-resolution radiography and routine histology. The fusion masses were graded and then statistically evaluated.
The results demonstrate a dose response of fusion mass scores to increasing current density. The highest current density (10 microA/cm) demonstrated a statistically higher fusion scores than the lowest currently density (0.83 microA/cm) at 6 weeks and statistically greater than both the middle (4 microA/cm) and lowest (0.83 microA/cm) current densities at 9 weeks. No differences were noted at 12 weeks, because all groups showed complete fusion.
This controlled study suggests that speed of fusion may be further improved by increasing the current density of the electrical stimulation.
电刺激已被证明可提高脊柱融合成功的几率。增加这些刺激器的电流密度可能会提高脊柱融合的速度和成功率。
本研究评估了不同电流密度对犬类脊柱融合速度和成功率的影响。
研究设计/环境:在12周的时间内,使用三种不同的电流密度(0.83微安/平方厘米、4微安/平方厘米和10微安/平方厘米)刺激犬类模型的脊柱融合。
通过影像学和组织学评估来确定每个时间段小关节融合的程度。
15只犬在L1-2和L4-5水平双侧进行脊柱小关节融合。每个融合部位使用三种电流密度之一进行电刺激。在6周、9周和12周时,使用高分辨率放射成像和常规组织学对标本进行评估。对融合块进行分级,然后进行统计学评估。
结果表明融合块评分对电流密度增加存在剂量反应。最高电流密度(10微安/平方厘米)在6周时显示出统计学上高于最低电流密度(0.83微安/平方厘米)的融合评分,在9周时统计学上高于中等电流密度(4微安/平方厘米)和最低电流密度(0.83微安/平方厘米)。在12周时未观察到差异,因为所有组均显示完全融合。
这项对照研究表明,通过增加电刺激电流密度可能会进一步提高融合速度。