Hubka M, Mantei K, Bolson E, Coady K, Sheehan F
University of Washington, Seattle, WA, USA.
Comput Cardiol. 2000;27:703-6.
Three-dimensional echocardiography (3D echo) by freehand scanning provides highly accurate measurements of left mass & volume using the piecewise smooth subdivision surface reconstruction method. The complexity of right ventricular (RV) geometry presents a challenge in accurate 3D assessments of its physical parameters. The authors measured the accuracy of their 3D echo technology in measuring RV mass and volume.
From 10 bovine hearts, the freewall mass and endocardial volume were measured. The hearts were imaged by 3D and the RV contours were manually traced. The endo- and epicardial surfaces were reconstructed. RV volume and mass from the reconstructions were compared to true measurements.
3D echo RV volume overestimated true volume (y=1.2x, r=0.998, SEE=3.2 ml, p<0.001), over a range of volumes (14-62 ml). The mean overestimation of RV volume by 3D echo was 6.2 ml, or 15.8% of the mean true volume. 3D echo mass agreed well with RV mass (y=1.0x, r=0.996, SEE=4.5 g, p=0.031) over a range of 47 to 79 g. Mean difference between 3D and true mass was 3.4 g, or 5.4% of the mean true mass.
RV mass and volume can be measured accurately from 3D echocardiograms acquired using freehand scanning and reconstruction by the piecewise smooth subdivision method.
通过徒手扫描的三维超声心动图(3D 超声)使用分段平滑细分曲面重建方法可对左心室质量和容积进行高度精确的测量。右心室(RV)几何形状的复杂性给其物理参数的准确三维评估带来了挑战。作者测量了他们的 3D 超声技术在测量 RV 质量和容积方面的准确性。
对 10 个牛心脏测量游离壁质量和心内膜容积。对心脏进行三维成像,并手动描绘 RV 轮廓。重建心内膜和心外膜表面。将重建得到的 RV 容积和质量与真实测量值进行比较。
在 14 - 62 ml 的容积范围内,3D 超声 RV 容积高估了真实容积(y = 1.2x,r = 0.998,标准误 = 3.2 ml,p < 0.001)。3D 超声对 RV 容积的平均高估为 6.2 ml,占平均真实容积的 15.8%。在 47 至 79 g 的范围内,3D 超声质量与 RV 质量吻合良好(y = 1.0x,r = 0.996,标准误 = 4.5 g,p = 0.031)。3D 超声与真实质量之间的平均差异为 3.4 g,占平均真实质量的 5.4%。
使用徒手扫描和分段平滑细分方法重建获得的三维超声心动图可准确测量 RV 质量和容积。