Hirai Toshinori, Korogi Yukunori, Ikushima Ichiro, Shigematsu Yoshinori, Morishita Shouji, Yamashita Yasuyuki
Department of Radiology, Kumamoto University School of Medicine, Japan.
Radiat Med. 2003 Sep-Oct;21(5):205-9.
Source images from three-dimensional time-of-flight MR angiography (3D TOF MRA) are useful for the diagnosis of cavernous dural arteriovenous fistulas. The purpose of this study was to assess the accuracy of source images from 3D TOF MRA compared with digital subtraction angiography (DSA) as a follow-up technique for the detection of residual or recurrent fistulas after treatment. We also assessed the effect of embolized materials in the evaluation of MRA.
In 14 patients who were treated with embolization and/or radiation therapy, follow-up MRA and DSA were performed after a mean follow-up duration of 3 months. Two observers independently reviewed the source images from MRA in regard to the presence of hyperintense regions within the cavernous sinus and enlargement of the draining veins. They also assessed whether embolized materials affected the evaluation.
Sensitivity, specificity, and positive predictive value of source images from MRA in revealing residual or recurrent fistulas were 100% (9 of 9 true positive findings), 80% (4 of 5 true negative findings), and 90% (9 of 10 positive findings), respectively. In all five patients treated with platinum coils, hypointense areas in the cavernous sinus were observed but did not affect image evaluation.
Source images from 3D TOF MRA provided accurate diagnosis with no effect of embolized materials in the follow-up of treated cavernous dural arteriovenous fistulas.
三维时间飞跃磁共振血管造影(3D TOF MRA)的源图像对海绵窦硬脑膜动静脉瘘的诊断很有用。本研究的目的是评估3D TOF MRA源图像与数字减影血管造影(DSA)相比作为治疗后检测残留或复发性瘘管的随访技术的准确性。我们还评估了栓塞材料对MRA评估的影响。
对14例接受栓塞和/或放射治疗的患者,平均随访3个月后进行随访MRA和DSA检查。两名观察者独立审查MRA的源图像,观察海绵窦内高强度区域的存在情况以及引流静脉的扩大情况。他们还评估了栓塞材料是否影响评估。
MRA源图像在显示残留或复发性瘘管方面的敏感性、特异性和阳性预测值分别为100%(9个真阳性结果中的9个)、80%(5个真阴性结果中的4个)和90%(10个阳性结果中的9个)。在所有5例接受铂线圈治疗的患者中,观察到海绵窦内有低强度区域,但不影响图像评估。
3D TOF MRA的源图像在治疗后的海绵窦硬脑膜动静脉瘘随访中提供了准确的诊断,且不受栓塞材料的影响。