Torn Margaret S, Davis Simon, Bird Jeffrey A, Shaw M Rebecca, Conrad Mark E
Lawrence Berkeley National Laboratory, Earth Sciences Division, ms 90-1116, One Cyclotron Rd, Berkeley, CA 94720, USA.
Rapid Commun Mass Spectrom. 2003;17(23):2675-82. doi: 10.1002/rcm.1246.
Stable carbon isotope ratios (13C/12C) are a valuable tool for studying a wide range of environmental processes, including carbon cycling and subsurface microbial activity. Recent advances in automated analysis provide the opportunity to increase greatly the ease and consistency of isotopic analysis. This study evaluated an automated headspace sampler linked to a commercially available CO2 preconcentration system and continuous flow isotope ratio mass spectrometer. Field sampling and analysis methods are illustrated for delta13C of soil respired CO2, from both tracer and natural abundance experiments, and dissolved inorganic carbon from contaminated groundwater. The automated system demonstrated accuracy, precision, and linearity, with standard errors below 0.1 per thousand for replicate gas standards run at concentrations varying five-fold. It measured 40 samples per 10-hour run, with concentrations ranging from ppb to percentage levels. In the field, gas samples were injected into nitrogen-filled autosampler vials, thereby allowing use of small sample volumes, control of analyte concentration, and direct analysis by the automated system with no further preparation. A significant linear relationship between standard concentrations and peak area allows for accurate estimates of sample CO2 concentration from the mass spectrometric data. The ability to analyze multiple small-volume samples with minimal off-line preparation should enhance the application of isotopes to well-replicated field experiments for process-level studies and spatial and temporal scaling.
稳定碳同位素比率(13C/12C)是研究广泛环境过程的宝贵工具,包括碳循环和地下微生物活动。自动分析技术的最新进展为大幅提高同位素分析的简便性和一致性提供了机会。本研究评估了一种与市售二氧化碳预浓缩系统和连续流动同位素比率质谱仪相连的自动顶空进样器。阐述了用于土壤呼吸二氧化碳δ13C(来自示踪剂和天然丰度实验)以及受污染地下水中溶解无机碳的现场采样和分析方法。该自动系统展示了准确性、精密度和线性度,对于浓度变化五倍的重复气体标准品,标准误差低于千分之一。每10小时运行可测量40个样品,浓度范围从十亿分比到百分比水平。在现场,气体样品被注入充氮的自动进样器小瓶中,从而可以使用小体积样品、控制分析物浓度,并由自动系统直接分析而无需进一步制备。标准浓度与峰面积之间的显著线性关系使得能够根据质谱数据准确估计样品二氧化碳浓度。以最少的离线制备分析多个小体积样品的能力应会增强同位素在用于过程水平研究以及空间和时间尺度的充分重复的现场实验中的应用。