Besch S, Snyder K V, Zhang P C, Sachs F
Huges Center for Single Molecule Biophysics, Physiology and Biophysical Sciences, SUNY at Buffalo, Buffalo, NY 14214, USA.
Cell Biochem Biophys. 2003;39(3):195-210. doi: 10.1385/cbb:39:3:195.
The Quesant Nomad atomic force microscope (AFM) was modified to produce a reliable patch-clamp AFM for demanding biologic applications. The AFM's laser optics forms the basis of a condenser that allows simultaneous Köhler illumination and AFM imaging on an inverted optical microscope. The original AFM scan head was replaced with plastic and glass to make it biologically inert. A bevel cut in the new scan head permits clearance for patch clamp pipets. Cantilevers are attached to the scan head with a quick setting silicone rubber that is readily removable. Software was developed to (a) automate a gentle approach and set a specific feedback force, (b) provide a mouse-driven control of the X-Y position of the probe tip and recall of saved locations, and (c) measure force-distance curves over user defined paths. Additional modifications were made to minimize mechanical noise. The patch-clamp AFM achieves 600 fA (3 kHz bandwidth) and 1 A RMS noise levels (10 kHz bandwidth). The correlation of electrical and mechanical information allows signal averaging and measures sub-Angstrom, sub-millisecond electromotile responses from cells.