Hummler Helmut D, Engelmann Anja, Pohlandt Frank, Högel Josef, Franz Axel R
Division of Neonatology and Pediatric Critical Care, Department of Pediatrics, Children's Hospital, University of Ulm, 89070 Ulm, Germany.
Intensive Care Med. 2004 Apr;30(4):709-13. doi: 10.1007/s00134-003-2116-1. Epub 2004 Jan 13.
To test the effects of low perfusion caused by emerging sepsis on the performance of two new pulse oximetry techniques: Masimo SET in comparison with Nellcor Oxismart XL.
Cohort study with random allocation of two pulse oximetry devices to two sensor sites. SETTING. University animal research facility.
Twenty-five adult, anesthetized, ventilated rabbits.
Pneumonia/sepsis was induced by tracheal instillation of E. coli.
Oxygen saturation was measured by pulse oximetry (SpO(2)) and recorded continuously until death. Arterial oxygen saturation (SaO(2)) was measured hourly by CO oximetry and whenever a difference of >5% between the devices occurred. SpO(2) sensors were positioned at both forelegs and switched hourly. There was no difference in total signal dropout time [median 3.8 min (range 0.4-66.6 min) vs 3.3 min (range 0-94.5 min), Masimo SET vs Oxismart XL]. There were fewer episodes with a false SpO(2) reading [1 (range 0-7) vs 2 (range 0-17)] using the Masimo SET vs the Oxismart XL as verified by CO oximetry; p<0.05. Average bias (SpO(2)-SaO(2)) was significantly different between the two devices, and variability of bias values increased across time with both devices.
Both devices were capable to measure SpO(2) during most of the experimental time in this model of low perfusion and therefore appear to be highly sensitive to pick up a signal; however, low perfusion caused by emerging sepsis may result in inaccurate measurements with both devices. These episodes were less common with the Masimo SET vs the Oxismart XL.
测试新发脓毒症导致的低灌注对两种新型脉搏血氧饱和度测定技术性能的影响:将Masimo SET与Nellcor Oxismart XL进行比较。
队列研究,将两种脉搏血氧饱和度测定设备随机分配至两个传感器部位。地点:大学动物研究设施。
25只成年、麻醉、机械通气的兔子。
通过气管内滴注大肠杆菌诱导肺炎/脓毒症。
通过脉搏血氧饱和度测定法测量氧饱和度(SpO₂),并持续记录直至死亡。每小时通过一氧化碳血氧饱和度测定法测量动脉血氧饱和度(SaO₂),并在两种设备之间出现>5%的差异时进行测量。SpO₂传感器置于双前肢,并每小时切换一次。总信号丢失时间无差异[中位数3.8分钟(范围0.4 - 66.6分钟)对比3.3分钟(范围0 - 94.5分钟),Masimo SET对比Oxismart XL]。经一氧化碳血氧饱和度测定法验证,使用Masimo SET时假SpO₂读数的情况少于Oxismart XL[1次(范围0 - 7次)对比2次(范围0 - 17次)];p<0.05。两种设备之间平均偏差(SpO₂ - SaO₂)显著不同,且随着时间推移,偏差值的变异性在两种设备中均增加。
在该低灌注模型的大部分实验时间内,两种设备均能够测量SpO₂,因此似乎对信号采集高度敏感;然而,新发脓毒症导致的低灌注可能导致两种设备测量不准确。与Oxismart XL相比,这些情况在Masimo SET中较少见。