Characterization of implantable microfabricated fluid delivery devices.

作者信息

Rathnasingham Ruben, Kipke Daryl R, Bledsoe Sanford C, McLaren John D

机构信息

Neural Engineering Laboratory, Biomedical Engineering, The University of Michigan, Ann Arbor, MI 48109-0506, USA.

出版信息

IEEE Trans Biomed Eng. 2004 Jan;51(1):138-45. doi: 10.1109/TBME.2003.820311.

Abstract

The formal characterization of the performance of microfluidic delivery devices is crucial for reliable in vivo application. A comprehensive laboratory technique was developed and used to optimize, calibrate and validate microfabricated fluid delivery devices. In vivo experiments were carried out to verify the accuracy and reliability of the pressure driven devices. Acute guinea pig experiments were conducted to measure the response to alpha-amino-3-hydroxy-5-methyl-4-isoxalone propionic acid, an excitatory neurotransmitter, at multiple locations in the inferior colliculus. A nondimensional parameter, Q, was successfully used to classify devices in terms of geometry alone (i.e., independent of fluid properties). Functional devices exhibited long-term linearity and reliability in delivering single phase, Newtonian fluids, in discrete volumes with a resolution of 500 picoliters at less than 0.45 lbf/in2 (30 mbar) pressure drop. Results for non-Newtonian fluids are not presented here. The acute results showed a proportional increase in the firing rate for delivered volumes of 2 nL up to 10 nL (at rates of between 0.1 and 1 nL/s). Flow characteristics are maintained during acute experiments and post-implant. A control experiment conducted with Ringer solution produced negligible effects, suggesting the results to be truly pharmacological. The experimental techniques employed have proven to be reliable and will be used for future calibration and testing of next generation chronic microfluidic delivery devices.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验