Continuous monitoring of right ventricular volume changes using a conductance catheter in the rabbit.

作者信息

Soldà P L, Pantaleo P, Perlini S, Calciati A, Finardi G, Pinsky M R, Bernardi L

机构信息

Department of Internal Medicine, University of Pavia, Italy.

出版信息

J Appl Physiol (1985). 1992 Nov;73(5):1770-5. doi: 10.1152/jappl.1992.73.5.1770.

Abstract

To assess the reliability of conductance (G) catheter for evaluating right ventricular (RV) volume changes, a miniature (3.5F) six-electrode catheter was developed and tested in 11 New Zealand rabbit hearts. In five animals the heart was excised; in six it was left in the thorax. RV conductance was recorded while the RV was filled with blood in 0.25-ml steps at different left ventricular (LV) volumes. Linear correlation of measured conductance vs. reference volumes was computed. RV conductance was highly correlated with reference volume [correlation coefficient (r) ranging from 0.991 to 0.999]. Slope of regression lines was not significantly affected by LV volume variations in 1-ml steps or by acute conductance changes of structures surrounding the heart, whereas the intercept was affected only by the 0- to 1-ml LV volume change. In four rabbits, RV conductance changes during a cardiac cycle [stroke volume- (SV) G] were compared in vivo with electromagnetic flow probe-derived estimates of SV (SVem) as stroke volume was varied by graded inferior vena caval occlusion. SV-G correlated well with SVem (r ranging from 0.92 to 0.96). This correlation persisted after the thorax was filled with saline; however, significant differences were found in individual slopes (P < 0.001). These results show that the conductance catheter has a potential to reliably monitor in vivo relative RV volume changes in small-animal hearts.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验