Daxboeck Florian, Dornbusch Hans Jürgen, Krause Robert, Assadian Ojan, Wenisch Christoph
Department of Pediatrics, University of Graz, Graz, Austria.
Diagn Microbiol Infect Dis. 2004 Jan;48(1):1-3. doi: 10.1016/j.diagmicrobio.2003.08.001.
A small but significant proportion of blood cultures processed by the BACTEC 9000 series systems is signaled positive, while subsequent Gram's stain and culture on solid media yield no pathogens. In this study, 15 "false-positive" vials (7 aerobes, 8 anaerobes) from 15 patients were investigated for the presence of bacteria and fungi by eubacterial 16S rDNA and panfungal 18S rDNA amplification, respectively. All samples turned out negative by both methods. Most patients (7) had neutropenia, which does not support the theory that high leukocyte counts enhance the generation of false-positive results. In conclusion, the results of this study indicate that false-negative results generated by the BACTEC 9000 series are inherent to the automated detection and not due to the growth of fastidious organisms.
BACTEC 9000系列系统处理的血液培养物中有一小部分虽显示阳性,但后续革兰氏染色及固体培养基培养却未发现病原体。在本研究中,对来自15名患者的15个“假阳性”培养瓶(7个需氧瓶,8个厌氧瓶)分别通过真细菌16S rDNA和泛真菌18S rDNA扩增来检测细菌和真菌的存在。两种方法检测所有样本结果均为阴性。大多数患者(7名)有中性粒细胞减少症,这并不支持白细胞计数高会增加假阳性结果的理论。总之,本研究结果表明,BACTEC 9000系列产生的假阴性结果是自动检测所固有的,而非由于苛养菌生长所致。