Comparison of modulation transfer function and through focus response with monofocal and bifocal IOLs.

作者信息

Knorz M C, Bedoya J H, Hsia T C, Neubert W J, Jones M, McCary B D, Seiberth V, Liesenhoff H

机构信息

Universitäts-Augenklinik, Klinikum Mannheim, Universität Heidelberg, Federal Republic of Germany.

出版信息

Ger J Ophthalmol. 1992;1(1):45-53.

PMID:1477618
Abstract

We analyzed modulation transfer function (MTF) and through focus response (TFR) of the TRUE VISTA bifocal IOL and monofocal IOLs and compared the experimental and clinical data. MTF and TFR were measured at different pupil sizes (2, 3, 4 mm), base powers (14, 21, 27 D), add powers (2, 3, 4 D), and distance zone diameters (1.25, 1.5, 1.75 mm). Standards were established to correlate MTF and contrast sensitivity, as well as TFR and the defocus curve measured clinically. MTF of TRUE VISTA decreased with increasing pupil size and base power. A near add of 4 D showed the best MTF at distant and near focus. MTF of TRUE VISTA at near focus was lower than at distant focus and MTF at distant focus was lower than MTF of monofocal IOLs. TFR and the defocus curve correlated closely. MTF and contrast sensitivity of TRUE VISTA at far and near focus also correlated closely. MTF and contrast sensitivity of monofocal IOLs and TRUE VISTA at distant focus did not correlate. MTF was considerably better with monofocal IOLs while contrast sensitivity was slightly better only at low contrast. However, MTF of monofocal IOLs was much higher than MTF of a normal eye while MTF of TRUE VISTA at distant focus was only slightly lower than the MTF of a normal eye. Using MTF of a normal eye as a threshold, our results demonstrated a close correlation between experimental and clinical findings. MTF and TFR may therefore reliably predict the performance of bifocal IOLs.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验