[Comparison of the properties of spot and ring electrodes in transesophageal cardiac pacing].

作者信息

Prochaczek F, Galecka J, Stípal R

机构信息

I. interní klinika Slezské lékarské fakulty Katowice.

出版信息

Cas Lek Cesk. 1992 Nov 20;131(23):700-2.

PMID:1477868
Abstract

The authors compare diagnostic oesophageal pacing of the left atrium by a ring and spot electrode using poles with an area of 20 mm2 oriented towards the left atrium. The examination was made in 15 volunteers, aged 37 +/- 5 years (range 28-43 years). In different depths of the oesophagus the threshold of stimulation was assessed (PS) and the threshold of perception of stimulation (PV) of the left atrium. At each of the examined levels the amplitude of waves A and V was determined as well as the A/V ratio. The change from a ring to a spot electrode reduces the PS from 12.2 +/- 2:3 (8.0-15.0 O) mA to 8.0 +/- 1.3 (6.0-10.0) mA. When the spot electrode is used, the difference between PS and PV is markedly reduced; one third of the examined subjects can be stimulated without subjective perception. The amplitude of the A wave recorded by means of the spot electrode was not reduced, as compared with the ring electrode and the optimal A/V ratio above 2:1 was recorded by the spot electrode on a longer section of the oesophagus. Evidence was also provided that the lowest PS is 3-4 cm beneath the site of recording of the maximum amplitude of wave A.

摘要

作者比较了使用面积为20平方毫米且朝向左心房的电极的环形电极和点状电极对左心房进行诊断性食管起搏的情况。对15名年龄在37±5岁(范围为28 - 43岁)的志愿者进行了检查。在食管的不同深度评估左心房的刺激阈值(PS)和刺激感知阈值(PV)。在每个检查水平测定A波和V波的振幅以及A/V比值。从环形电极改为点状电极可使PS从12.2±2.3(8.0 - 15.0)毫安降至8.0±1.3(6.0 - 10.0)毫安。使用点状电极时,PS与PV之间的差异明显减小;三分之一的受检者在无主观感知的情况下即可受到刺激。与环形电极相比,点状电极记录的A波振幅未降低,并且点状电极在食管更长的节段上记录到了高于2:1的最佳A/V比值。还证实最低PS位于记录到A波最大振幅部位下方3 - 4厘米处。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验