Mastragelopulos N, Sarkar M R, Kaissling G, Bähr R, Daub D
Chirurgische Klinik, Abteilung Allgemein- und Thoraxchirurgie, Städtischen Klinikums Karlsruhe.
Chirurg. 1992 Dec;63(12):1053-4.
The report is about argon embolism in a 37-year-old female patient during laparoscopic cholecystectomy and coagulation of a liver bed bleeding with the "Argon Beam One". The presented case report is to show that in spite of technical perfection and correct use of the Argon Beam One coagulator unfavorable anatomical conditions can cause life threatening gas embolism.
该报告讲述了一名37岁女性患者在腹腔镜胆囊切除术期间发生氩气栓塞的情况,以及使用“氩束一号”对肝床出血进行凝血的过程。所呈现的病例报告旨在表明,尽管氩束一号凝血器技术完善且使用正确,但不利的解剖条件仍可能导致危及生命的气体栓塞。