[Thermoluminescent dosimetry for radiation exposure of the breast during film and xeromammography (author's transl)].

作者信息

Säbel M, Ruff A, Weishaar J

出版信息

Rofo. 1978 May;128(5):616-22. doi: 10.1055/s-0029-1230915.

Abstract

Measurements of radiation exposure of the breast during various mammographic procedures were carried out with LiF dosimeters. Using non-screen film (Kodak Definix Medical) at 30 kv., average skin dose in 105 patients was 10.1 R per exposure. During xeroradiography (45 kv.) skin dose was about 25% of this and for a film-screen system (Mammoray RP 3/MR 50 at 30 kv.) only 7% of this value. Skin dose (at constant potential) rises almost exponentially with increasing tissue thickness. The age of the patient and the type of tissue in the breast has a significant effect on the radiation dose. From the measurements, an average dose per exposure was calculated for the entire breast and for the glandular tissue. This gave a result of 1.6 rd (16 mGy) and 1.0 rd (10 mGy).

摘要

使用氟化锂剂量计对各种乳房X线摄影检查过程中的乳房辐射暴露进行了测量。在30千伏电压下使用无屏胶片(柯达Definix Medical)时,105名患者每次曝光的平均皮肤剂量为10.1伦琴。在干板X线摄影(45千伏)时,皮肤剂量约为该值的25%,而对于胶片-增感屏系统(30千伏下的Mammoray RP 3/MR 50),仅为该值的7%。(在恒定电势下)皮肤剂量几乎随组织厚度的增加呈指数上升。患者的年龄和乳房组织类型对辐射剂量有显著影响。根据测量结果,计算出了整个乳房和腺体组织每次曝光的平均剂量。结果分别为1.6拉德(16毫戈瑞)和1.0拉德(10毫戈瑞)。