[Changes in the lumbar disk following use of non-automated percutaneous discectomy. A biomechanical study].

作者信息

Castro W H, Jerosch J, Brinckmann P

机构信息

Klinik und Poliklinik für Allgemeine Orthopädie, Westfälischen Wilhelms-Universität Münster.

出版信息

Z Orthop Ihre Grenzgeb. 1992 Nov-Dec;130(6):472-8. doi: 10.1055/s-2008-1039655.

Abstract

For a biomechanical study 8 specimens of the lumbar spine (L1-S1) were scanned with MRI. After that 21 motion segments were prepared for the experiments. The paraspinal musculature as well as the posterior elements were removed. From each disc 1,0 g, 2,0 g, and 3,0 g of nuclear material were consecutively removed under an axial load of 500 N. At the beginning of every standardized experiment of a motion segment, as well as after the excision of every gram of nuclear material, intradiscal pressure, the height of the disc and the contour of the disc were determined with an axial load on the motion segment of 1000 N. The change in radial bulging of the disc was measured using the computer-assisted recordings of each disc contour. All the data were statistically analyzed using the Friedman-test. The nuclear material removed from the discs ('in-vitro nuclear material') was placed in isotonic saline over 45 minutes and again weighted ('wet weight'). Af that the nuclear material was freeze-dried ('dry weight'). The results of this study showed that after the excision of every gram of nuclear material the height of the disc decreased in average 0,77 mm. Also intradiscal pressure reduces significantly: from the initial stage, on average, 9, 19 bar to the endstage (= after the removal of 3,0 g), on average, 2, 58 bar. Radial bulge increases with every gram removed, on average 0,23 mm. The weight of the wet nuclear material is 2,6 times as much as that of the in-vitro nuclear material and 11 times as much as that of the freeze-dried material.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验