Courtney S E, Weber K R, Spohn W A, Bender C V, Malin S W, Gotshall R W
Division of Newborn Medicine, Wright State University, Dayton, Ohio.
Pediatr Pulmonol. 1992 Jun;13(2):113-6. doi: 10.1002/ppul.1950130210.
To examine the effects of varying inspiratory/expiratory ratio (I/E) on cardiorespiratory function during high-frequency oscillation (HFO), 11 saline-lavaged rabbits were ventilated at I/E = 1:2, 1:1.5, 1.5:1, and 2:1 in a paired comparison to a baseline of I/E = 1:1. HFO was delivered by a SensorMedics model 3100 oscillator at a frequency of 10 Hz. Pressure amplitude and proximal mean airway pressure (PPaw) were held constant as I/E was varied from baseline to the experimental I/E. During each paired observation, PaO2, PaCO2, cardiac output, blood pressure, and distal mean airway pressure (DPaw) were measured. We found that as I/E was increased or decreased from 1:1, no significant changes in PaO2, PaCO2, blood pressure, or cardiac output occurred. We conclude that in this model, varying I/E has no significant effect on oxygenation, ventilation, or cardiovascular function.
为研究高频振荡(HFO)期间不同吸气/呼气比(I/E)对心肺功能的影响,对11只经盐水灌洗的家兔在I/E = 1:2、1:1.5、1.5:1和2:1的条件下进行通气,并与I/E = 1:1的基线进行配对比较。通过SensorMedics 3100型振荡器以10 Hz的频率进行高频振荡。当I/E从基线变化到实验性I/E时,压力振幅和近端平均气道压(PPaw)保持恒定。在每对观察期间,测量动脉血氧分压(PaO2)、动脉血二氧化碳分压(PaCO2)、心输出量、血压和远端平均气道压(DPaw)。我们发现,当I/E从1:1增加或减少时,PaO2、PaCO2、血压或心输出量均无显著变化。我们得出结论,在该模型中,改变I/E对氧合、通气或心血管功能无显著影响。