使用实时复合B型超声对颈动脉斑块进行特征分析。

Characterization of carotid artery plaques using real-time compound B-mode ultrasound.

作者信息

Kern Rolf, Szabo Kristina, Hennerici Michael, Meairs Stephen

机构信息

Department of Neurology, Universitätsklinikum Mannheim, University of Heidelberg, 68135 Mannheim, Germany.

出版信息

Stroke. 2004 Apr;35(4):870-5. doi: 10.1161/01.STR.0000120728.72958.4A. Epub 2004 Mar 4.

Abstract

BACKGROUND AND PURPOSE

Real-time compound ultrasound imaging is a new technique for improving the image quality of B-mode scanning. We investigated the value of this method for the characterization of atherosclerotic plaques in the internal carotid artery.

METHODS

Thirty-two patients (22 men, 10 women; mean age, 75 years) with plaques of the internal carotid artery as identified by high-resolution B-mode scanning were investigated with real-time compound ultrasound imaging with the use of a 5- to 12-MHz dynamic range linear transducer on a duplex scanner. Two independent observers rated plaque morphology according to a standardized protocol.

RESULTS

The majority of plaques was classified as predominantly echogenic and as plaques of irregular surface, whereas ulcerated plaques were rarely observed. The interobserver agreement for plaque surface characterization was good for both compound ultrasound (kappa=0.72) and conventional B-mode (kappa=0.65). For the determination of plaque echogenicity, the reproducibility of compound ultrasound (kappa(w)=0.83) was even higher than that of conventional B-mode ultrasound (kappa(w)=0.74). According to a semiquantitative analysis, real-time compound ultrasound was rated superior in the categories plaque texture resolution, plaque surface definition, and vessel wall demarcation. Furthermore, there was a significant reduction of acoustic shadowing and reverberations.

CONCLUSIONS

Real-time compound ultrasound is a suitable technique for the characterization of atherosclerotic plaques, showing good general agreement with high-resolution B-mode imaging. This advanced technique allows reduction of ultrasound artifacts and improves the assessment of plaque texture and surface for enhanced evaluation of carotid plaque morphology.

摘要

背景与目的

实时复合超声成像是一种提高B型扫描图像质量的新技术。我们研究了该方法在颈动脉粥样硬化斑块特征描述方面的价值。

方法

对32例经高分辨率B型扫描确诊为颈动脉斑块的患者(22例男性,10例女性;平均年龄75岁),使用双功扫描仪上的5至12MHz动态范围线性换能器进行实时复合超声成像检查。两名独立观察者根据标准化方案对斑块形态进行评分。

结果

大多数斑块被分类为以强回声为主且表面不规则的斑块,而溃疡斑块很少见。复合超声(kappa=0.72)和传统B型超声(kappa=0.65)在斑块表面特征描述方面观察者间一致性良好。对于斑块回声性的判定,复合超声(kappa(w)=0.83)的可重复性甚至高于传统B型超声(kappa(w)=0.74)。根据半定量分析,实时复合超声在斑块纹理分辨率、斑块表面清晰度和血管壁分界方面评分更高。此外,声影和混响明显减少。

结论

实时复合超声是一种适用于描述动脉粥样硬化斑块特征的技术,与高分辨率B型成像总体一致性良好。这种先进技术可减少超声伪像,改善对斑块纹理和表面的评估,以增强对颈动脉斑块形态的评价。