Naruse Shigeta, Mori Kazuhiko, Kojo Mina, Hieda Osamu, Kinoshita Shigeru
Department of Ophthalmology, Kyoto Prefectural University of Medicine, Kyoto, Japan.
J Cataract Refract Surg. 2004 Feb;30(2):390-7. doi: 10.1016/S0886-3350(03)00614-X.
To evaluate the applicability of pressure phosphene tonometry in measuring intraocular pressure (IOP) after laser in situ keratomileusis (LASIK).
Baptist Eye Clinic, Kyoto, Japan.
Thirty-six eyes of 22 consecutive patients who had had LASIK for myopia were enrolled in this prospective comparative study. The mean age of the patients was 32.6 years +/- 9.0 (SD). The IOP was measured using the pressure phosphene tonometer (FPT), Goldmann applanation tonometry (GAT), and noncontact tonometry (NCT). The patients themselves took measurements with the FPT before and 1 week after surgery. One-way analysis of variance followed by a Bonferroni/Dunn multiple comparison post hoc test was administered to compare the preoperative and postoperative measurements with each tonometer. These data were represented graphically using a method described by Bland and Altman. The relationship between each tonometric change and the central corneal thickness (CCT) or keratometry (K) changes after LASIK were evaluated with by Pearson correlation coefficient.
Statistically significant differences were found between preoperative and postoperative measurements with GAT and NCT but not with FPT. Preoperative and postoperative FPT readings conformed to the British Standard for reproducibility of a standard test method for IOP estimation. No significant correlation between FPT changes and CCT or K changes was found.
Pressure phosphene tonometry measures IOP through the upper eyelid, and changes in corneal shape do not appear to have an effect on the IOP. The IOP measurements with FPT were constant before and after LASIK.
评估压平式眼压计在准分子原位角膜磨镶术(LASIK)后测量眼压(IOP)的适用性。
日本京都浸信会眼科诊所。
本前瞻性对照研究纳入了22例连续接受LASIK治疗近视的患者的36只眼。患者的平均年龄为32.6岁±9.0(标准差)。使用压平式眼压计(FPT)、Goldmann压平眼压计(GAT)和非接触眼压计(NCT)测量眼压。患者在手术前和术后1周自行使用FPT进行测量。采用单因素方差分析,随后进行Bonferroni/Dunn多重比较事后检验,以比较每种眼压计术前和术后的测量结果。这些数据采用Bland和Altman描述的方法进行图形表示。通过Pearson相关系数评估LASIK后每种眼压测量变化与中央角膜厚度(CCT)或角膜曲率(K)变化之间的关系。
GAT和NCT术前和术后测量结果存在统计学显著差异,但FPT没有。术前和术后FPT读数符合英国眼压估计标准测试方法再现性标准。未发现FPT变化与CCT或K变化之间存在显著相关性。
压平式眼压计通过上眼睑测量眼压,角膜形状的变化似乎对眼压没有影响。LASIK术前和术后使用FPT测量的眼压是恒定的。